9.5.2 Behavior

The Target Resource shall be rendered as specified in Section 8.3.5.1.

A success response shall contain the Target Resource in an Acceptable Rendered Media Type. See Section 8.7.4.

9.5.2.1 Frame Number

If this Query Parameter is supported and is present in the request,the origin server shall use the specified frame as the Target Resource.

However, if any of the following are true:

  • the Target Resource is not a multi-frame image or video,

  • the number of parameter values is not equal to one, or

  • the parameter value is not a positive integer less than or equal to the number of frames in the Instance

the origin server shall return a 400 (Bad Request) responseand may include a payload containing an appropriate error message.