| DICOM PS3.18 2024e - Web Services |
|---|
The "viewpointup" parameter specifies the vertical orientation of the camera within the Viewpoint Coordinate System (VCS). See Section C.11.30.1 in PS3.3.
The syntax of this parameter is:
%s"viewpointup =" ux "," uy "," uz
ux, uy and uz |
Viewpoint up direction (i.e., the direction that the top of the camera is pointing to). A vector (x,y,z) in the VCS. |
This corresponds to the Viewpoint Up Direction (0070,1605) attribute. See Section C.11.30 in PS3.3.
If the Target Resource is a Volumetric Presentation Stateand any orientation Query Parameters are present, the origin server shall apply the query parameter(s) instead of the geometry attributes in the Multi-Planar Reconstruction Geometry Module, or the Volume Render Geometry Module.
| DICOM PS3.18 2024e - Web Services |
|---|