The "orientation" parameter specifies the patient's orientation as seen by the camera for the current 3D or MPR Volumetric Presentation View.
The syntax of this parameter is:
%s"orientation =" 1#( %s"a" / %s"p" / %s"r" / %s"l" / %s"h" / %s"f" )
a |
|
p |
|
r |
|
l |
|
h |
The camera is viewing the patient's head (i.e., from above): |
f |
The camera is viewing the patient's feet (i.e., from below): |
These values correspond to the differently capitalized values of the Patient Orientation (0020,0020) and Image Orientation (Patient) (0020,0037). See Section C.7.6.1.1.1 in PS3.3 and Annex A in PS3.17.
If the Target Resource is a Volumetric Rendering Presentation State and any orientation Query Parameters are present, the origin server shall apply the query parameter(s) instead of the geometry attributes in the Multi-Planar Reconstruction Geometry Module, or the Volume Render Geometry Module.
This is intended to allow the user to adjust orientation after viewing the initial orientation defined in the Volumetric Presentation State.
Orientation is used to select the desired standard anatomic position of the rendered volume. For example, "a" specifies the direction of the rendered volume that most closely aligns with the patient's anterior plane. Viewing angles beyond the standard orthogonal anatomic positions are controlled by camera orientation parameters (i.e., "viewpointposition", "viewpointlookat", or "viewpointup").
The origin server shall determine a Viewpoint Position (0070,1603), a Viewpoint LookAt Point (0070,1604) and a Viewpoint Up Direction (0070,1605) based on the value of the "orientation" parameter.
If both the "orientation" parameter and any of the camera orientation parameters are present, the origin server shall return a 400 (Bad Request) and may include an appropriate Status Report.