The "renderingmethod" parameter specifies the display algorithm to be applied to the Volume Data.
The syntax of this parameter is:
%s"renderingmethod=" 1#( %s"volume_rendered" / %s"maximum_ip" / %s"minimum_ip" / %s"average_ip" )
These values correspond to the differently capitalized values of Rendering Method (0070,120D). See Section C.11.23 in PS3.3 and Section C.11.30 in PS3.3.
There is no parameter to control the type of projection used during rendering. Rendered 3D Volume Resources use Orthographic projection. See Figure C.11.30-1 in PS3.3.
For Rendered MPR Volume Resources, this parameter describes the display algorithm to apply when the slab thickness is greater than one voxel. This parameter value is typically average_ip.
If "renderingmethod" is not present, the origin server may apply a default rendering method, based on the resource, or alternatively, return 400 (Bad Request) and may include an appropriate Status Report.
If the Target Resource is a Volumetric Presentation State, the origin server shall return a 400 (Bad Request) and may include an appropriate Status Report.