A success response shall contain a list of matching results in an Acceptable Media Type. See Section 8.7.4.
A failure response payload may contain a Status Report describing any failures, warnings, or other useful information.
For each matching Study, the origin server response shall contain Attributes in accordance with Table 10.6.3-3. The "Type" column in the table below refers to the Query/Retrieve Attribute Types defined in Section C.2.2.1 “Attribute Types” in PS3.4. The unique key for a Study resource Search response is the Study Instance UID (0020,000D).
Table 10.6.3-3. Study Resource Search Response Payload
While some of the above Attributes are optional in Table C.6-1 “Patient Level Attributes for the Patient Root Query/Retrieve Information Model” in PS3.4, Table C.6-2 “Study Level Keys for the Patient Root Query/Retrieve Information Model” in PS3.4, and Table C.6-5 “Study Level Keys for the Study Root Query/Retrieve Information Model” in PS3.4, they are consistent with those required in [IHE RAD TF Vol2] Table 4.14-1.
If either Prior Record Key (0008,041C) or Record Key (0008,041B) is present in the request and the Origin Server supports these attributes, then the responses are ordered as defined in Section C.6.4 “Repository Query SOP Class” in PS3.4.
In addition, the response shall contain:
All other Study level Attributes passed as match or includefield parameters in the request that are supported by the origin server.
If the includefield parameter has been specified in the request, and its value is "all", all Study Level Attributes specified as Optional Keys in Section C.6.1.1.2 “Patient Level” in PS3.4, Section C.6.1.1.3 “Study Level” in PS3.4, and Section C.6.2.1.2 “Study Level” in PS3.4 that are supported by the origin server.
If a Private Data Element has been specified as an include parameter and it is supported by the origin server, the Private Data Element and its corresponding Private Creator Element.
Series or Instance Level Attributes contained in includefield parameters shall not be returned.