| DICOM PS3.18 2024e - Web Services |
|---|
For each matching Instance, the origin server shall return all Attributes listed in Table 10.6.3-5, if present in the Instance. The Type column in the table below refers to the Query/Retrieve Attribute Types defined in Section C.2.2.1 “Attribute Types” in PS3.4. The unique key for an Instance resource search response is the SOP Instance UID (0008,0018).
Table 10.6.3-5. Instance Resources Search Response Payload
While some of the above Attributes in are optional in Table C.6-4 “Composite Object Instance Level Keys for the Patient Root Query/Retrieve Information Model” in PS3.4, they are consistent with the those required in [IHE RAD TF Vol2] Table 4.14-1.
If either Prior Record Key (0008,041C) or Record Key (0008,041B) is present in the request and the origin server supports these attributes, then the responses are ordered as defined in Section C.6.4 “Repository Query SOP Class” in PS3.4.
In addition, the response shall contain:
All other Instance Level Attributes passed as match parameters, or Study, Series, or Instance Level Attributes passed as includefield parameters that are supported by the origin server.
If the "includefield" parameter has been specified in the request and its value is "all", Attribute include all Instance Level Attributes specified as Optional Keys in Section C.6.1.1.5 “Composite Object Instance Level” in PS3.4 that are supported by the origin server.
If the Target Resource is All Instances, then include all Study level Attributes specified in Section 10.6.3.3.1.
If the Target Resource is All Instances or Study's Instances, then include all Series level Attributes specified in Section 10.6.3.3.2.
If a Private Data Element has been specified as an include parameter and it is supported by the origin server, the Private Data Element and its corresponding Private Creator Element.
The response may optionally include:
the Available Transfer Syntax UID (0008,3002) to describe the Transfer Syntaxes that the origin server can assure will be supported for retrieval of the SOP Instance. See Section C.6.1.1.5.2 in PS3.4 .
| DICOM PS3.18 2024e - Web Services |
|---|