C.3.1 SOP Classes and Transfer Syntaxes
Application Profiles in this class, STD-US, shall support the appropriate Information Object Definitions (IODs) and Transfer Syntaxes in the following table. In the role of FSU or FSC the application can choose one of the three possible Transfer Syntaxes to create a SOP Instance. In the role of FSR an application shall support all Transfer Syntaxes defined for the respective STD-US Application Profile.
Table C.3-1. Ultrasound SOP Classes and Transfer Syntaxes
|
Information Object Definition
|
SOP Class UID
|
Transfer Syntax Name
|
Transfer Syntax UID
|
|
Basic Directory
|
1.2.840.10008.1.3.10
|
Explicit VR Little Endian
|
1.2.840.10008.1.2.1
|
|
Ultrasound Image
|
1.2.840.10008.5.1.4.1.1.6.1
|
Explicit VR Little Endian
|
1.2.840.10008.1.2.1
|
|
Ultrasound Image
|
1.2.840.10008.5.1.4.1.1.6.1
|
RLE Lossless
|
1.2.840.10008.1.2.5
|
|
Ultrasound Image
|
1.2.840.10008.5.1.4.1.1.6.1
|
JPEG Baseline (Process 1)
|
1.2.840.10008.1.2.4.50
|
|
Ultrasound Multi-frame Image
|
1.2.840.10008.5.1.4.1.1.3.1
|
Explicit VR Little Endian
|
1.2.840.10008.1.2.1
|
|
Ultrasound Multi-frame Image
|
1.2.840.10008.5.1.4.1.1.3.1
|
RLE Lossless
|
1.2.840.10008.1.2.5
|
|
Ultrasound Multi-frame Image
|
1.2.840.10008.5.1.4.1.1.3.1
|
JPEG Baseline (Process 1)
|
1.2.840.10008.1.2.4.50
|
C.3.1.1 Ultrasound Single and Multi-frame Pixel Formats Supported
The Application Profiles of the STD-US class require that all ultrasound image objects only be stored using the values described in the US Image Module in PS3.3 and the specializations used for the Ultrasound Single and Multi-Frame IODs.
In the role of FSU or FSC the application can choose any of the supported Photometric Interpretations described in the US Image Module in PS3.3 to create an IOD.
In the role of FSR, an application shall support all Photometric Interpretations described in the US Image Module in PS3.3.
Table C.3-2 describes restrictions on the use of various Transfer Syntaxes with the supported Photometric Interpretations for both single and Multi-frame Images.
Table C.3-2. Defined Photometric Interpretation and Transfer Syntax Pairs
|
Photometric Interpretation Value
|
Transfer Syntax
|
Transfer Syntax UID
|
|
MONOCHROME2
|
Explicit VR Little Endian
RLE Lossless
|
1.2.840.10008.1.2.1
1.2.840.10008.1.2.5
|
|
RGB
|
Explicit VR Little Endian
RLE Lossless
|
1.2.840.10008.1.2.1
1.2.840.10008.1.2.5
|
|
PALETTE COLOR
|
Explicit VR Little Endian
RLE Lossless
|
1.2.840.10008.1.2.1
1.2.840.10008.1.2.5
|
|
YBR_FULL
|
RLE Lossless
|
1.2.840.10008.1.2.5
|
|
YBR_FULL_422
|
Explicit VR Little Endian
JPEG Baseline (Process 1)
|
1.2.840.10008.1.2.1
1.2.840.10008.1.2.4.50
|