| DICOM PS3.7 2024e - Message Exchange |
|---|
The C-MOVE-RSP Message contains fields as defined in Table 9.3-10 and in Annex C. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the C-MOVE service definition unless otherwise noted in Table 9.3-10. Fields not specified in the C-MOVE service definition but present in Table 9.3-10 are required by the DIMSE-C protocol.
Table 9.3-10. C-MOVE-RSP Message Fields
|
The even number of bytes from the end of the value field to the beginning of the next group. |
||||
|
This field distinguishes the DIMSE-C operation conveyed by this Message. The value of this field shall be set to 8021H for the C-MOVE-RSP Message. |
||||
|
Shall be set to the value of the Message ID (0000,0110) field used in associated C-MOVE Message. |
||||
|
This field indicates if a Data Set is present in the Message. This field shall be set to the value of 0101H (Null) if no Data Set is present; any other value indicates a Data Set is included in the Message. |
||||
|
The value of this field depends upon the status type. Annex C defines the encoding of the status types defined in the service definition. |
||||
|
The number of remaining sub-operations to be invoked for this C-MOVE operation. |
||||
|
The number of C-STORE sub-operations invoked by this C-MOVE operation that have completed successfully. |
||||
|
The number of C-STORE sub-operations invoked by this C-MOVE operation that have failed. |
||||
|
The number of C-STORE sub-operations invoked by this C-MOVE operation that generated warning responses. |
||||
|
A Data Set that encodes attributes providing status information about the C-MOVE operation. See Section 9.1.4.1.6. |
| DICOM PS3.7 2024e - Message Exchange |
|---|