| DICOM PS3.7 2024e - Message Exchange |
|---|
Table E.1-1. Command Fields
|
The even number of bytes from the end of the value field to the beginning of the next group. |
|||||
|
This field distinguishes the DIMSE operation conveyed by this Message. This field shall be set to one of the following values: |
|||||
|
Implementation-specific value that distinguishes this Message from other Messages. |
|||||
|
Shall be set to the value of the Message ID (0000,0110) field used in associated request Message. |
|||||
|
Shall be set to the DICOM AE Title of the destination DICOM AE to which the C-STORE sub-operations are being performed. |
|||||
|
This field indicates if a Data Set is present in the Message. This field shall be set to the value of 0101H if no Data Set is present; any other value indicates a Data Set is included in the Message. |
|||||
|
Confirmation status of the operation. See Annex C. |
|||||
|
If status is Cxxx, then this field contains a list of the elements in which the error was detected. |
|||||
|
This field contains an application-specific text description of the error detected. |
|||||
|
This field shall optionally contain an application-specific error code. |
|||||
|
Contains the UID of the SOP Instance for which this operation occurred. |
|||||
|
Contains the UID of the SOP Instance for which this operation occurred. |
|||||
|
This field contains an Attribute Tag for each of the n Attributes applicable. |
|||||
|
The number of remaining C-STORE sub-operations to be invoked for the operation. |
|||||
|
The number of C-STORE sub-operations associated with this operation that have completed successfully. |
|||||
|
The number of C-STORE sub-operations associated with this operation that have failed. |
|||||
|
The number of C-STORE sub-operations associated with this operation that generated warning responses. |
|||||
|
Contains the DICOM AE Title of the DICOM AE that invoked the C-MOVE operation from which this C-STORE sub-operation is being performed. |
|||||
|
Contains the Message ID (0000,0110) of the C-MOVE-RQ Message from which this C-STORE sub-operation is being performed. |
| DICOM PS3.7 2024e - Message Exchange |
|---|