| DICOM PS3.4 2024e - Service Class Specifications |
|---|
An implementation may conform to one of the Inventory Query/Retrieve Service SOP Classes as an SCU, SCP or both. The Conformance Statement shall be in the format defined in PS3.2.
An implementation that conforms to the Inventory FIND SOP Class as an SCU shall support queries against the Inventory Q/R Information Model using the C-FIND SCU Behavior described for the Basic Worklist Management Service Class (see Section K.4.1.2 and Section JJ.3.1).
An implementation that conforms to the Inventory MOVE SOP Class as an SCU shall support transfers against the Inventory Q/R Information Model using the C-MOVE SCU baseline behavior described for the Query/Retrieve Service Class (see Section C.4.2.2.1 and Section JJ.3.2).
An implementation that conforms to the Inventory GET SOP Class as an SCU shall support transfers against the Inventory Q/R Information Model using the C-GET SCU baseline behavior described for the Query/Retrieve Service Class (see Section C.4.3.2.1 and Section JJ.3.3).
| DICOM PS3.4 2024e - Service Class Specifications |
|---|