| DICOM PS3.4 2024e - Service Class Specifications |
|---|
An implementation that conforms to one of the SOP Classes of the Media Storage Service Class:
shall meet the requirements specified in Section I.2.3;
shall meet the requirements specified in PS3.10;
shall perform M-READ Operations according to the SOP Class specification identified by the SOP Class UID in the Meta File Information. M-READ of unsupported SOP Classes shall simply result in ignoring such stored Data Sets;
shall perform M-WRITE Operations according to the SOP Class specification identified by the SOP Class UID in the Meta File Information;
shall support the Media Storage Directory SOP Class (stored in the DICOMDIR File). Directories containing a Directory Information Module shall be updated by an FSU. Directories containing no Directory Information Module shall not be updated by an FSU;
shall read DICOMDIR Files without a Directory Information Module or with a Directory Information Module including Directory Records of a Type not supported by the implementation.
| DICOM PS3.4 2024e - Service Class Specifications |
|---|