The DIMSE Service shown in Table EE.2.2-1 is applicable to the Display System IOD under the Display System SOP Class.
This section describes the behavior of the DIMSE services that are specific for this IOD. The general behavior of the DIMSE services is specified in PS3.7.
N-GET is used to retrieve information from an instance of the Display System SOP Class.
To reduce the size and complexity of EE.2.2.1-2, a macro notation is used.
Table EE.2.2.1-1. Table Result Context Macro
The Attributes that may be retrieved are shown in Table EE.2.2.1-2.
Table EE.2.2.1-2. Display System N-GET Attributes
|
(Required if an extended or replacement character set is used) |
||
|
(Required if Institution Code Sequence (0008,0082) is not present. May be present otherwise.) |
||
|
(Required if Institution Name (0008,0080) is not present.May be present otherwise.) |
||
|
(Required if the value of Display Function Type (0028,7019) is GAMMA) |
||
|
(Required if the value of Display Function Type (0028,7019) is USER_DEFINED) |
||
|
(Required if the value of Display Function Type (0028,7019) is USER_DEFINED) |
||
|
(Required if the value of Display Function Type (0028,7019) is USER_DEFINED. May be present otherwise.) |
||
|
(Required if Reflected Ambient Light (2010,0160) is present.) |
||
|
>>>>Include Table EE.2.2.1-1 “Table Result Context Macro” |
||
|
>>>>Include Table EE.2.2.1-1 “Table Result Context Macro” |
||
|
>>>>>>Include Table CC.2.5-2a “UPS Code Sequence Macro” |
||
|
(Required if Test Pattern Code Sequence (0028,702C) is absent in this item.May be present otherwise.) |
||
|
(Required if the Referenced SOP Instance is a multi-frame image and the reference does not apply to all frames, and Referenced Segment Number (0062,000B) is not present.) |
||
|
(Required if the Referenced SOP Instance is a Segmentation or Surface Segmentation and the reference does not apply to all segments and Referenced Frame Number (0008,1160) is not present.) |
||
|
>>>>>Include Table CC.2.5-2a “UPS Code Sequence Macro” |
||
|
>>>>Include Table EE.2.2.1-1 “Table Result Context Macro” |
||
|
>>>>>Include Table CC.2.5-2a “UPS Code Sequence Macro” |
||
|
(Required if the value of White Point Flag (0028,7021) is YES.) |
||
|
(Required if Reflected Ambient Light (2010,0160) is present.) |
||
|
>>>>Include Table EE.2.2.1-1 “Table Result Context Macro” |
||
|
(Required if Reflected Ambient Light (2010,0160) is present.) |
||
The SCU uses the N-GET to request the SCP to provide the contents of a Display System SOP Instance. The SCU shall specify in the N-GET request primitive the UID of the SOP Instance from which Attributes are to be returned.
The SCU shall specify the list of Display System Attributes for which values are to be returned. The SCU shall not specify Attributes which are defined within a Sequence, but rather specify the sequence itself to be returned in its entirety.
The SCU shall specify in the N-GET request primitive the well-known UID of the SOP Instance.
The SCP shall return the values for the specified Attributes of the Display System SOP Instance.
The SCP shall return the Status Code for the requested SOP Instance retrieval. The meaning of success, warning, and Failure Status Codes are defined in PS3.7.