DICOM PS3.3 2024e - Information Object Definitions

F.5.48 Inventory Directory Record Definition

The Directory Record is based on the specification of Section F.3. It is identified by a Directory Record Type (0004,1430) of Value "INVENTORY". Table F.5-48 lists the set of Keys with their associated Types for such a Directory Record Type. The description of these Keys may be found in the Inventory IE of the Inventory IOD. This Directory Record shall be used to reference an Inventory SOP Instance. This type of Directory Record may reference a Lower-Level Directory Entity that includes one or more Directory Records as defined in Table F.4-1.

Table F.5-48. Inventory Keys

Key

Tag

Type

Attribute Description

Specific Character Set

(0008,0005)

1C

Required if an extended or replacement character set is used in one of the Keys.

Content Date

(0008,0023)

1

Content Time

(0008,0033)

1

Inventory Purpose

(0008,0401)

2

Inventory Level

(0008,0403)

1

Include Table C.38.2-1 “Scope of Inventory Macro Attributes”

Inventory Completion Status

(0008,0426)

1

Total Number of Study Records

(0008,0428)

1


DICOM PS3.3 2024e - Information Object Definitions