The following restrictions apply when a Compensator is represented as a Surface Segmentation SOP Instance:
Segment Sequence (0062,0002) shall contain only one item.
Segmented Property Category Code Sequence (0062,0003) shall contain (260787004, SCT, "Physical object").
Segmented Property Type Code Sequence (0062,000F) shall contain (130340, DCM, "Physical Compensator").
Finite Volume (0066,000E) shall be YES.
Manifold (0066,0010) shall be YES.
Point Coordinates Data (0066,0016) shall be defined in mm in the IEC GANTRY coordinate system where the origin (0,0,0) is at the isocenter. This means that for a single sided compensator, the z-coordinate for the vertices at the base of the compensator has the value of Isocenter to Compensator Tray Distance (300A,02E4).