| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.8-59. RT Treatment Summary Record Module Attributes
|
Status of the Treatment at the time the Treatment Summary was created. See Section C.8.8.23.1. |
|||
|
Sequence describing current state of planned vs. delivered fraction groups. |
|||
|
References Fraction Group Number (300A,0071) in Fraction Group Sequence (300A,0070) in the referenced RT Plan. |
|||
|
Number of fractions delivered as of Treatment Summary Report. |
|||
|
Uniquely identifies Dose Reference specified by Dose Reference Number (300A,0012) in Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan referenced in Referenced RT Plan Sequence (300C,0002) of RT General Treatment Record Module. |
|||
|
Uniquely identifies Dose Reference specified by Dose Reference Number (300A,0012) in Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan referenced in Referenced RT Plan Sequence (300C,0002) of RT General Treatment Record Module. |
|||
The RT Treatment Summary Record IOD may contain references to related RT Treatment Session Record IODs. These references are contained within the Referenced Treatment Record Sequence (3008,0030) of the RT General Treatment Record Module.
Enumerated Values:
Patient is currently not undergoing treatment, but a resumption date is known
Patient is currently not undergoing treatment, but resumption of treatment is planned at an unknown date
Patient has stopped treatment without completing the planned course
A change in the Current Treatment Status (or any other field) in a RT Treatment Summary Record Object shall define a new Instance of the RT Treatment Summary Record IOD.
| DICOM PS3.3 2024e - Information Object Definitions |
|---|