Table C.8-58. RT Brachy Session Record Module Attributes
|
Total number of treatments (Fractions) planned for current Fraction Group. |
|||
|
Type of brachytherapy treatment technique. See RT Plan IOD. |
|||
|
Identification number of the Source. The value of Source Number (300A,0212) shall be unique within the Recorded Source Sequence (3008,0100) in which it is created. |
|||
|
User-supplied identifier for the radioactive source model that was used for the source in the treatment plan of which this session record is based to. See Section C.8.8.15.15. |
|||
|
Measurement unit of Source Strength. Required if the source is not a gamma-emitting (photon) source. May be present otherwise. |
|||
|
Air Kerma Rate in air of Isotope specified at Source Strength Reference Date (300A,022C) and Source Strength Reference Time (300A,022E) (in µGy h-1 at 1 m). Value shall be zero for non-gamma sources. |
|||
|
Source Strength of Isotope at Source Strength Reference Date (300A,022C) and Source Strength Reference Time (300A,022E), in units specified in Source Strength Units (300A,0229). Required if the source is not a gamma-emitting (photon) source. See Section C.8.8.15.13. |
|||
|
Reference date for Reference Air Kerma Rate (300A,022A) or Source Strength (300A,022B) of Isotope. |
|||
|
Reference time for Air Kerma Rate (300A,022A) or Source Strength (300A,022B) of Isotope. |
|||
|
Sequence of Application Setups for RT Treatment Record for current RT Plan. |
|||
|
References application setup specified by Application Setup Number (300A,0234) in Application Setup Sequence (300A,0230) in RT Brachy Application Setups Module within referenced RT Plan. |
|||
|
Results of check-wire travel through all channels of current Application Setup. |
|||
|
Sequence of verification images obtained during delivery of current Application Setup. One or more Items are permitted in this Sequence. See Note 1. |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Total Reference Air Kerma for current Application Setup, i.e., the sum of the products of the Air Kerma Rates of each Source in each Channel with its respective Channel Time (µGy at 1 m). Value shall be zero for non-gamma sources. |
|||
|
Sequence of doses measured during treatment delivery, summed over entire session. |
|||
|
Uniquely references Dose Reference specified by Dose Reference Number (300A,0012) in Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan. Required if Referenced Measured Dose Reference Number (3008,0082) is not present. It shall not be present otherwise. |
|||
|
Uniquely references Measured Dose Reference specified by Measured Dose Reference Number (3008,0064) in Measured Dose Reference Sequence (3008,0010). Required if Referenced Dose Reference Number (300C,0051) is not present. It shall not be present otherwise. |
|||
|
Measured Dose in units specified by Dose Units (3004,0002) in Sequence referenced by Measured Dose Reference Sequence (3008,0010) or Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan as defined above. |
|||
|
Uniquely identifies Dose Reference specified by Dose Reference Number (300A,0012) in Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan. Required if Referenced Calculated Dose Reference Number (3008,0092) is not present. It shall not be present otherwise. |
|||
|
Uniquely identifies Calculated Dose Reference specified by Calculated Dose Reference Number (3008,0072) within Calculated Dose Reference Sequence (3008,0070). Required if Referenced Dose Reference Number (300C,0051) is not present. It shall not be present otherwise. |
|||
|
Identifies the reason why the current treatment beam has terminated. Expected if the Treatment Termination Status (3008,002A) is not NORMAL. One or more items are permitted in this Sequence. See 2. |
|||
|
Machine-specific termination codes. Expected if the RT Treatment Termination Reason Code is (110501, DCM, "Equipment failure"). One or more Items are permitted in this Sequence. See 2. |
|||
|
A user-readable description for an abnormal termination. See 2. |
|||
|
Conditions under which treatment was verified by a verification system. |
|||
|
Sequence of Brachy Accessory Devices associated with current Application Setup. |
|||
|
Identification number of the Brachy Accessory Device. The value of Brachy Accessory Device Number (300A,0262) shall be unique within the Application Setup in which it is created. |
|||
|
User or machine supplied identifier for Brachy Accessory Device. |
|||
|
Identification number of the Channel. The value of Channel Number (300A,0282) shall be unique within the Application Setup in which it is created. |
|||
|
The channel to be delivered, specified by the value of Channel Number (300A,0282) in referenced RT Plan. |
|||
|
Length of Channel (mm). See RT Plan IOD. |
|||
|
Length of Channel (in mm) defined as the distance between the connector on the afterloader and the center of the distal-most possible position of the source. See Section C.8.8.15.16. |
|||
|
The total physical inner length of channel (in mm). Specifies the distance between the connector on the afterloader and the end of the channel. Required if Channel Effective Length (300A,0271) is present. |
|||
|
Identification of the Channel connection on the afterloader. |
|||
|
Total amount of time in seconds, scaled for the current source delivery strength and other delivery factors, specified to be delivered at the time of treatment between the first Control Point and the final Control Point for the current Channel. In the case of resuming a partially delivered treatment, the Specified Channel Total time will only include the remainder to be treated. See Section C.8.8.22.2. |
|||
|
Total amount of time in seconds actually delivered between Control Point 0 and final Control Point of the Brachy Control Point Delivered Sequence (3008,0160) for current Channel. |
|||
|
Number of Pulses specified per fraction for current Channel. Required if Brachy Treatment Type (300A,0202) is PDR. See Section C.8.8.22.1. |
|||
|
Number of Pulses actually delivered per fraction for current Channel. Required if Brachy Treatment Type (300A,0202) is PDR. See Section C.8.8.22.1. |
|||
|
Pulse repetition interval (sec) specified for current Channel. Required if Brachy Treatment Type (300A,0202) is PDR. See Section C.8.8.22.1 |
|||
|
Pulse repetition interval (sec) actually delivered for current Channel. Required if Brachy Treatment Type (300A,0202) is PDR. See Section C.8.8.22.1. |
|||
|
Sequence of doses measured during treatment delivery, summed over entire session. |
|||
|
Uniquely references Dose Reference specified by Dose Reference Number (300A,0012) in Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan. Required if Referenced Measured Dose Reference Number (3008,0082) is not present. It shall not be present otherwise. |
|||
|
References Measured Dose Reference specified by Measured Dose Reference Number (3008,0064) in Measured Dose Reference Sequence (3008,0010). Required if Referenced Dose Reference Number (300C,0051) is not present. It shall not be present otherwise. |
|||
|
Uniquely identifies Dose Reference specified by Dose Reference Number (300A,0012) in Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan. Required if Referenced Calculated Dose Reference Number (3008,0092) is not present. It shall not be present otherwise. |
|||
|
Uniquely identifies Calculated Dose Reference specified by Calculated Dose Reference Number (3008,0072) within Calculated Dose Reference Sequence (3008,0070). Required if Referenced Dose Reference Number (300C,0051) is not present. It shall not be present otherwise. |
|||
|
Identification number of the Source Applicator. The value of Source Applicator Number (300A,0290) shall be unique within the Channel in which it is created. |
|||
|
Length of Source Applicator (mm), defined as the distance between the connector of the applicator and the distal-most position of the source. |
|||
|
Length of Source Applicator Tip (in mm), defined as the distance between the outer tip of the applicator and the center of the distal-most possible position of the source. Required if Channel Effective Length (300A,0271) is present. See Section C.8.8.15.16. |
|||
|
Distance of path along channel (mm) between adjacent (potential) dwell positions. Required if Source Movement Type (300A,0288) is STEPWISE. |
|||
|
Identification number of the Transfer Tube. The value of Transfer Tube Number (300A,02A2) shall be unique within the Channel in which it is created. |
|||
|
Length of Transfer Tube of current afterloading Channel (mm). Required if value Transfer Tube Number (300A,02A2) is not zero length. |
|||
|
Sequence of Channel Shields associated with current Channel. One or more Items are permitted in this Sequence. See RT Plan IOD for description of Channel Shields. |
|||
|
Identification number of the Channel Shield. The value of Channel Shield Number (300A,02B2) shall be unique within the Channel in which it is created. |
|||
|
Uniquely identifies the referenced Source within the Recorded Source Sequence (3008,0100) for current Application Setup. |
|||
|
Date on which the source(s) exited the safe. Required if Recorded Channel Sequence (3008,0130) is present and Brachy Treatment Type (300A,0202) is not MANUAL or PDR. |
|||
|
Time on which the source(s) exited the safe. Required if Recorded Channel Sequence (3008,0130) is present and Brachy Treatment Type (300A,0202) is not MANUAL or PDR. |
|||
|
Date on which the source(s) returned to the safe. Required if Recorded Channel Sequence (3008,0130) is present and Brachy Treatment Type (300A,0202) is not MANUAL or PDR. |
|||
|
Time on which the source(s) returned to the safe. Required if Recorded Channel Sequence (3008,0130) is present and Brachy Treatment Type (300A,0202) is not MANUAL or PDR. |
|||
|
Number of control points in Channel. For an N-segment Channel there will be 2N (stepwise movement) or N+1 (continuous movement) control points. |
|||
|
Sequence of machine configurations describing this Channel. The number of Items in this Sequence shall equal the value of Number of Control Points (300A,0110). See RT Plan IOD and Section C.8.8.22.1 for description of Brachy Control Point Delivered Sequence. |
|||
|
Uniquely identifies Control Point specified by Control Point Index (300A,0112) within Application Setup referenced by Referenced Brachy Application Setup Number (300C,000C). |
|||
|
Date when the delivery of radiation at this control point began. For the final control point this shall be the Date when the previous control point ended. |
|||
|
Time when the delivery of radiation at this control point began. For the final control point this shall be the Time when the previous control point ended. |
|||
|
Distance in mm between current Control Point Position and the center of the distal-most possible Source position in current Channel. See RT Plan IOD. |
|||
|
Sequence of parameters that were overridden during the administration of the treatment immediately prior to the current control point. |
|||
|
Contains the Data Element Tag of the Attribute that was overridden. |
|||
|
Identification of the operator who authorized override. Only a single Item is permitted in this Sequence. |
|||
|
>>>>>Include Table 10-1 “Person Identification Macro Attributes” |
|||
|
User-defined description of reason for override of parameter specified by Override Parameter Pointer (3008,0062). |
|||
|
Brachy Control Point Delivered Sequence for each PDR treatment pulse. Number of Items in the Sequence shall be equal to the Delivered Number of Pulses (3008,0138). |
|||
|
Identification Number of this delivered Pulse. The pulse numbers for a treatment start at 1 and increase monotonically by 1. A given SOP Instance might only contain some of the pulses of the given treatment. |
|||
|
List of control points for this pulse. See Section C.8.8.22.1. |
|||
|
Uniquely identifies Control Point specified by Control Point Index (300A,0112) within Application Setup referenced by Referenced Brachy Application Setup Number (300C,000C). |
|||
|
Date when the delivery of radiation at this control point began. For the final control point, this shall be the Date when the previous control point ended. |
|||
|
Time when the delivery of radiation at this control point began. For the final control point, this shall be the Time when the previous control point ended. |
|||
|
Distance in mm between current Control Point Position and the center of the distal-most possible Source position in current Channel. See Section C.8.8.15.9. |
|||
|
Parameters which were overridden during the administration of the treatment immediately prior to the current control point. |
|||
|
Identification of the operator who authorized override. Only a single Item is permitted in this Sequence. |
|||
|
>>>>>>Include Table 10-1 “Person Identification Macro Attributes” |
|||
The Referenced Verification Image Sequence (300C,0040) may contain either images taken specifically for verification of the brachy application setup or reference images used in place of verification images, as might be done in HDR treatment planning.
Treatment Termination Code (3008,002B) was previously included in this Module but has been retired. See PS3.3-2023a. The RT Treatment Termination Reason Code Sequence (300A,0715) and Machine-Specific Treatment Termination Code Sequence (300A,0716) should be used for machine readable codes and Treatment Termination Description (300A,0730) for human readable text respectively.
In Brachytherapy treatment techniques where Brachy Treatment Type (300A,0202) is PDR, the Brachy Control Point Sequence (300A,02D0) shall consist of 2N Items, where N = Delivered Number of Pulses (3008,0138). Each control point pair shall represent a single pulse and consist of the first control point of the pulse and the last control point of the pulse.
The Brachy Pulse Control Point Delivered Sequence (3008,0173) is consistent with the Brachy Control Point Delivered Sequence (3008,0160) as defined in the enclosing Recorded Channel Sequence (3008,0130) except all control points delivered are listed in this Sequence as opposed to listing only the start and end control points in the Brachy Control Point Delivered Sequence (3008,0160).