| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.8-47. RT Tolerance Tables Module Attributes
|
Sequence of tolerance tables to be used for delivery of treatment plan. One or more Items are permitted in this Sequence. See Note 1. |
|||
|
Identification number of the Tolerance Table. The value of Tolerance Table Number (300A,0042) shall be unique within the RT Plan in which it is created. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Gantry Angle. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Gantry Pitch Angle. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Beam Limiting Device Angle. |
|||
|
Sequence of beam limiting device (collimator) tolerances. One or more Items are permitted in this Sequence. |
|||
|
Type of beam limiting device (collimator). Enumerated Values: |
|||
|
Maximum permitted difference (in mm) between planned and delivered leaf (element) or jaw positions for current beam limiting device (collimator). |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Patient Support Angle. |
|||
|
Maximum permitted difference (in degrees) between planned and delivered Table Top Eccentric Angle. |
|||
|
Maximum permitted difference (in degrees) between the planned and delivered Table Top Pitch Angle. |
|||
|
Maximum permitted difference (in degrees) between the planned and delivered Table Top Roll Angle. |
|||
|
Maximum permitted difference (in mm) between planned and delivered Table Top Vertical Position. |
|||
|
Maximum permitted difference (in mm) between planned and delivered Table Top Longitudinal Position. |
|||
|
Maximum permitted difference (in mm) between planned and delivered Table Top Lateral Position. |
| DICOM PS3.3 2024e - Information Object Definitions |
|---|