| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.8-10 specifies the Attributes of the NM Isotope Module, which describe the isotope administered for the acquisition.
Table C.8-10. NM Isotope Module Attributes
|
Sequence of Items that describe the energy window groups used. Zero or more Items shall be included in this Sequence. The number of Items shall be equal to Number of Energy Windows (0054,0011). The first Item corresponds to frames with value of 1 in the Energy Window Vector (0054,0010), the second Item with value 2, etc. |
|||
|
The lower limit of the energy window in KeV. See Section C.8.4.10.1.1 for further explanation. |
|||
|
The upper limit of the energy window in KeV. See Section C.8.4.10.1.2 for further explanation. |
|||
|
BCID 18 “Radiopharmaceutical Isotope”. Code Meaning (0008,0104) shall be Type 3 for historical reasons. |
|||
|
Sequence that identifies the administration route for the radiopharmaceutical. |
|||
|
BCID 11 “Administration Route”. Code Meaning (0008,0104) shall be Type 3 for historical reasons. |
|||
|
Time of start of injection. See Section C.8.4.10.1.5 for further explanation. |
|||
|
Time of end of injection. See Section C.8.4.10.1.6 for further explanation. |
|||
|
Total amount of radionuclide injected. See Section C.8.4.10.1.7 for further explanation. |
|||
|
Unique identification of the administration of the radiopharmaceutical to the patient. |
|||
|
The Item number in the Energy Window Information Sequence to which the following calibration data relates. The Items are numbered starting from 1. |
|||
|
Pre-injection syringe count rate in counts/sec. See Section C.8.4.10.1.8 for further explanation. |
|||
|
Post-injection residue syringe count rate in counts/sec. See Section C.8.4.10.1.9 for further explanation. |
|||
|
BCID 25 “Radiopharmaceutical”. Code Meaning (0008,0104) shall be Type 3 for historical reasons. |
|||
|
Sequence of Items that describes the intervention drugs used. One or more Items are permitted in this Sequence. |
|||
|
BCID 10 “Interventional Drug”. Code Meaning (0008,0104) shall be Type 3 for historical reasons. |
|||
|
Sequence that identifies the administration route for the intervention drug. |
|||
|
BCID 11 “Administration Route”. Code Meaning (0008,0104) shall be Type 3 for historical reasons. |
|||
|
Time of administration of the intervention drug, using the same time base as for Acquisition Time (0008,0032). |
|||
|
Time of completion of administration of the intervention drug, using the same time base as for Acquisition Time (0008,0032). |
|||
Energy Window Lower Limit (0054,0014) is the acquisition energy window lower limit in KeV for acceptance of scintillation events into this Isotope.
Energy Window Upper Limit (0054,0015) is the acquisition energy window upper limit in KeV for acceptance of scintillation events into this Isotope.
Radiopharmaceutical Start Time (0018,1072) is the actual time of radiopharmaceutical administration to the patient for imaging purposes, using the same time base as for Acquisition Time (0008,0032).
Radiopharmaceutical Stop Time (0018,1073) is the actual ending time of radiopharmaceutical administration to the patient for imaging purposes, using the same time base as for Acquisition Time (0008,0032).
Radionuclide Total Dose (0018,1074) is the radiopharmaceutical dose administered to the patient measured in MegaBecquerels (Mbq) at the Radiopharmaceutical Start Time.
Syringe Counts (0018,1045) is the pre-injection syringe acquisition count rate measured in counts/sec, corrected to Acquisition Time (0008,0032) if necessary.
| DICOM PS3.3 2024e - Information Object Definitions |
|---|