| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.8.26.5-1 defines the Attributes used to interpret the test results of a patient's visual field test.
Table C.8.26.5-1. Visual Field Static Perimetry Test Results Module Attributes
|
Average sensitivity of the test points of the visual field, in dB. Required if Content Item Modifier Sequence (0040,0441) within the Performed Protocol Code Sequence (0040,0260) contains an Item with the value (261004008, SCT, "Diagnostic"). May be present otherwise. |
|||
|
Information that represents the statistically normal results for patients from a referenced data base. Only a single Item shall be included in this Sequence. Required if Visual Field Test Normals Flag (0024,0063) is YES. |
|||
|
>Include Table 10-22 “Externally-Sourced Data Set Identification Macro Attributes” |
|||
|
Weighted average deviation from the age corrected normal field, in dB. |
|||
|
Probability value and software algorithm used to provide the normality for the global deviation. Only a single Item shall be included in this Sequence. Required if Global Deviation Probability Normals Flag (0024,0059) is YES. |
|||
|
The percentile of the Global Deviation from Normal (0024,0066) value within the normal population, in percent.0024. |
|||
|
>>Include Table 10-19 “Algorithm Identification Macro Attributes” |
|||
|
Probability value and software algorithm used to provide the normality for the local deviation. Only a single Item shall be included in this Sequence. Required if Local Deviation Probability Normals Flag (0024,0072) is YES. |
|||
|
The percentile of the Localized Deviation From Normal (0024,0068) value within the normal population, in percent. |
|||
|
>>Include Table 10-19 “Algorithm Identification Macro Attributes” |
|||
|
Average deviation of sensitivity for the repeated test locations, in dB. This is used to determine the consistency of the patient's responses. Required if Short Term Fluctuation Calculated (0024,0074) is YES. |
|||
|
Whether the short term fluctuation probability was calculated. |
|||
|
The percentile of the Short Term Fluctuation (0024,0075) value within the normal population, in percent. Required if Short Term Fluctuation Probability Calculated (0024,0076) is YES. |
|||
|
Whether the corrected localized deviation from normal was calculated. |
|||
|
Weighted square root of loss variance corrected for short term fluctuation, in dB. Required if Corrected Localized Deviation From Normal Calculated (0024,0078) is YES. |
|||
|
Corrected Localized Deviation From Normal Probability Calculated |
Whether the corrected localized deviation from Normal probability was calculated. |
||
|
The percentile of the Corrected Localized Deviation From Normal (0024,0079) value within the normal population, in percent. Required if Corrected Localized Deviation From Normal Probability Calculated (0024,0080) is YES. |
|||
|
Information about various visual field indexes related to test results. |
|||
|
>Include Table C.8.26.3-2 “Ophthalmic Visual Field Global Index Macro Attributes” |
|||