| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.8.25.16-1 defines the Attributes used to calculate the power of an intraocular lens to be placed in a patient's eye. This is typically done for both eyes for comparison, but may be for just one eye. Values used may be measurements acquired using the axial length measurements device, manually entered, or derived from a referenced SOP class or other data source.
Table C.8.25.16-1. Intraocular Lens Calculations Module Attributes
|
Calculations of intraocular lens power for a patient's right eye. One or more Items shall be included in this Sequence. Required if the device calculated intraocular lens power for the right eye. |
|||
|
>Include Table C.8.25.16-2 “Intraocular Lens Calculations Macro Attributes” |
|||
|
Calculations of intraocular lens power for a patient's left eye. One or more Items shall be included in this Sequence. Required if the device calculated intraocular lens power for the left eye. |
|||
|
>Include Table C.8.25.16-2 “Intraocular Lens Calculations Macro Attributes” |
|||
| DICOM PS3.3 2024e - Information Object Definitions |
|---|