Table C.8.25.14-4 specifies the Attributes of the Ophthalmic Axial Measurements Related Information Macro.
Table C.8.25.14-4. Ophthalmic Axial Measurements Related Information Macro Attributes
|
Related information about an axial length measurement being performed on an ultrasound device. Only a single Item shall be included in this Sequence. Required if Ophthalmic Axial Measurements Device Type (0022,1009) is ULTRASOUND. |
|||
|
The sound velocity used in calculating the Ophthalmic Axial Length Measurement Value (0022,1019), in m/sec. |
|||
|
Method used to select which acquired axial length measurement values are kept. See Section C.8.25.14.1.1.5 for further explanation. |
|||
|
Source of the value recorded in Ophthalmic Axial Length (0022,1019). |
|||
|
DCID 4240 “Ophthalmic Measurement or Calculation Data Source” |
|||
|
A free text description of the axial length measurement data source. |
|||
|
Related information about an axial length measurement being performed on an optical device. Only a single Item shall be included in this Sequence. Required if Ophthalmic Axial Measurements Device Type (0022,1009) is OPTICAL. |
|||
|
The ratio of signal power to the background noise power (Psignal / Pnoise). |
|||
|
The source of the value in Ophthalmic Axial Length (0022,1019). |
|||
|
DCID 4240 “Ophthalmic Measurement or Calculation Data Source” |
|||
|
A free text description of the axial length measurement data source. |
|||