| DICOM PS3.3 2024e - Information Object Definitions |
|---|
The following sections contain Functional Group Macros specific to the X-Ray 3D IOD.
The Attribute descriptions in the Functional Group Macros are written as if they were applicable to a single frame (i.e., the Macro is part of the Per-Frame Functional Groups Sequence). If an Attribute is applicable to all frames (i.e., the Macro is part of the Shared Functional Groups Sequence) the phrase "this frame" in the Attribute description shall be interpreted to mean " for all frames".".
Table C.8.21.5.1-1 specifies the Attributes of the X-Ray 3D Frame Type Macro.
Table C.8.21.5.1-1. X-Ray 3D Frame Type Macro Attributes
|
A Sequence that describes general characteristics of this frame. |
|||
|
Type of Frame. A multi-valued Attribute analogous to Image Type (0008,0008). Enumerated Values and Defined Terms are the same as those for the four values of Image Type (0008,0008), except that the value MIXED is not allowed. See Section C.8.21.1.1.1. For Breast Tomosynthesis Image See Section C.8.21.6.1.1. |
|||
|
>Include Table C.8-131 “Common CT/MR and Photoacoustic Image Description Macro Attributes” |
|||
|
The Item number of the X-Ray 3D Reconstruction Sequence (0018,9530) that describes the characteristics of the 3D Reconstruction to which this frame is part of. Required if the X-Ray 3D Reconstruction Sequence (0018,9530) is present. |
|||
| DICOM PS3.3 2024e - Information Object Definitions |
|---|