| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.7-12b specifies the Attributes of the Enhanced Contrast/Bolus Module, which describe the contrast/bolus used in the acquisition of the Image.
Table C.7-12b. Enhanced Contrast/Bolus Module Attributes
|
Sequence that identifies one or more contrast agents administered prior to or during the acquisition. |
|||
|
Identifying number, unique within this SOP Instance, of the agent administered. Used to reference this particular agent from the Contrast/Bolus Functional Group Macro. The number shall be 1 for the first Item and increase by 1 for each subsequent Item. |
|||
|
Sequence that identifies the route of administration of contrast agent. |
|||
|
Total volume administered in milliliters of diluted contrast agent. |
|||
|
Percentage by volume of active ingredient in the total volume. |
|||
|
Absorption of the ingredient greater than the absorption of water (tissue). See Section C.7.6.4b.1. |
|||
|
T1 Relaxivity of the MR Contrast/Bolus used specified in s-1*mmol-1 specified at body temperature in human blood plasma. |
|||
|
Sequence that describes one or more phases of contrast administered. |
|||
|
Volume administered during this phase in milliliters of diluted contrast agent. |
|||
|
Rate of administration in milliliters/sec. Only a single value shall be present. |
|||
|
Duration of injection in seconds. Only a single value shall be present. |
|||
Contrast/Bolus Ingredient Opaque (0018,9425) specifies the type of relative X-Ray absorption of the contrast/bolus ingredient, compared to the X-Ray absorption of water.
Enumerated Values:
The Contrast/Bolus Ingredient Opaque (0018,9425) Attribute determines the sign of the gradient of X-Ray beam intensity from inside to outside the injected vessel, thus allowing optimal settings of the image processing applications (e.g., vessel edge detection, etc.), see Figure C.7.6.4b-1.
The relative gray level of the injected vessel with respect to the gray level of the water of Pixel Data (7FE0,0010) is determined by the Contrast/Bolus Ingredient Opaque (0018,9425) and by the Pixel Intensity Relationship Sign (0028,1041). For example, if the contrast/bolus ingredient is more radio graphically dense than water (i.e., YES), and the Pixel Intensity Relationship Sign (0028,1041) is -1, then the contrast/bolus ingredient is represented by higher values of Pixel Data than water.