The RT Projection Imaging Request Geometry Macro describes the location of the imaging source and image receptor to be used for acquisition.
Table C.36.2.4.1-1. RT Projection Imaging Request Geometry Macro Attributes
|
The method of specifying the location and orientation of the imaging source. Enumerated Values:
|
|||
|
Parameters describing the location of the Imaging Source and the Image Receptor by the means of matrices. Required if Imaging Source Location Specification Type (3002,0111) is present and has a value of ABSOLUTE_MATRIX. |
|||
|
>Include Table C.36.2.4.2-1 “Matrix-based RT Imaging Geometry Macro Attributes” |
|||
|
Parameters describing the location and orientation of the image receptor by the means of parameters. Required if Imaging Source Location Specification Type (3002,0111) is present and has a value of ABSOLUTE_PARAMS or RELATIVE_PARAMS. |
|||
|
Index of the RT Control Point of the RT Radiation SOP Instance referenced in the Referenced Baseline Parameters RT Radiation Instance Sequence (3002,011E) which defines the values against which the relative values have to be applied. Required if Imaging Source Location Specification Type (3002,0111) has a value of RELATIVE_PARAMS. |
|||
|
>Include Table C.36.2.4.3-1 “Parameterized RT Imaging Geometry Macro Attributes” |
|||
If Imaging Source Location Specification Type (3002,0111) has a value of ABSOLUTE_PARAMS the parameters have to be applied as is.
If Imaging Source Location Specification Type (3002,0111) has a value of RELATIVE_PARAMS the parameters represent delta values between the parameters of the referenced RT Control Point in Radiation SOP Instance referenced in the Referenced Baseline Parameters RT Radiation Instance Sequence (3002,011E). Only those parameters whose value is not zero shall be included in the Imaging Device Location Parameter Sequence (3002,0113).