| DICOM PS3.3 2024e - Information Object Definitions |
|---|
The MV Radiation Image Acquisition Parameters Macro contains parameters specifying details of X-Ray generation for image acquisition using MV-level radiation.
Table C.36.2.4.7-1. MV Radiation Image Acquisition Parameters Macro Attributes
|
How the value of imaging energy is derived from the device configuration. Required if Radiation Generation Mode Sequence (300A,067B) is not present. |
|||
|
The beam parameters of the imaging energy. Required if Energy Derivation Code Sequence (3002,0133) is not present. |
|||
|
>Include Table C.36.2.2.7-1 “Radiation Generation Mode Macro Attributes” |
DCID 9525 “Radiation Therapy Particle” for Radiation Type Code Sequence (300A,067F). DCID 9521 “Radiotherapy Treatment Energy Unit” for Energy Unit Code Sequence (300A,0684). DCID 9549 “Radiation Generation Mode Type” for Radiation Fluence Modifier Code Sequence (300A,0683). The Number of Radiation Generation Modes (300A,0685) shall have the value one. |
||
|
The maximum allowed exposure described in Meterset values between start and stop of acquisition. The radiation shall be stopped at latest when the difference between the start value and the current value exceeds this value. The unit is defined in the Radiation Dosimeter Unit Sequence (300A,0658). |
|||
|
Measurement unit of the machine dosimeter. Required if Maximum Cumulative Meterset Exposure (3002,0134) is present. |
|||
|
The nominal rate of delivery of the Meterset during acquisition. Required if the acquisition is not taken simultaneously during the delivery of the therapeutic radiation. The unit is defined in the Delivery Rate Unit Sequence (300A,063E). |
|||
|
The unit of a delivery rate value. Required if Delivery Rate (300A,063D) is present and has a value. |
|||
| DICOM PS3.3 2024e - Information Object Definitions |
|---|