Table C.17.6-2 specifies the Attributes of a Key Object Selection Document. These Attributes identify and provide context for the Key Object Selection Document.
Table C.17.6-2. Key Object Document Module Attributes
|
Identifies Requested Procedures to which this Document pertains. One or more Items shall be included in this Sequence. Required if this Document pertains to at least one Requested Procedure. |
|||
|
>Include Table C.17-3c “Referenced Request Macro Attributes” |
|||
|
List of all Composite SOP Instances referenced in the Content Sequence (0040,A730). The current evidence is considered to be only the set of Instances referenced within the Key Object Selection. |
|||
|
>Include Table C.17-3 “Hierarchical SOP Instance Reference Macro Attributes” |
|||
|
Duplicates of this document, stored with different SOP Instance UIDs. One or more Items shall be included in this Sequence. Required if this Key Object Selection document references Instances in more than one Study. See Section C.17.2.2 and Section C.17.6.2.1 for further explanation and conditions. |
|||
|
>Include Table C.17-3 “Hierarchical SOP Instance Reference Macro Attributes” |
|||
If the Current Requested Procedure Evidence Sequence (0040,A375) references SOP Instances both in the current Study and in one or more other Studies, this document shall be duplicated into each of those other Studies, and the duplicates shall be referenced in the Identical Documents Sequence (0040,A525).