| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.11.21-1 specifies the Attributes of the XA/XRF Presentation State Presentation Module.
Table C.11.21-1. XA/XRF Presentation State Presentation Module Attributes
|
Describes for one or more SOP Instances the recommended playback and display preferences. |
|||
|
Sequence of Items where each Item provides reference to a selected set of Image SOP Class/SOP Instance pairs that are defined in the Presentation State Relationship Module. One or more Items shall be included in this Sequence. Required if display sequences in this Item do not apply to all the images listed in the Presentation State Relationship Module. |
|||
|
>>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Describes the preferred playback sequencing for a Multi-frame Image. |
|||
|
Sequence that specifies the display frame rate of a selected set of frames. The Items are ordered in increasing frame number. The range of the frames may not overlap and the ranges shall be adjacent. |
|||
|
The Frame Number of the first frame of the set of frames to be displayed in this Item. |
|||
|
The Frame Number of the last frame of the set of frames to be displayed in this Item. |
|||
|
A flag indicating that the range of frames in this Item may be skipped. |
|||
|
Recommended rate at which the frames of this Item should be displayed in frames/second. |
|||
|
Edge enhancement filter percentage that is recommended by the pixel data creator as filter presetting for display purposes. The value of 100 corresponds to the maximum filter strength that can be applied by a specific application displaying the image. |
|||
|
The percentage of visibility of the mask frame during a subtracted display. A value of 0 corresponds to subtracted display, a value of 100 corresponds to un-subtracted display (native). See Section C.8.19.7.1. Required if Recommended Viewing Mode (0028,1090) equals SUB. |
|||