| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table C.11.12-1 specifies the Attributes of the Presentation State Shutter Module, which specialize Attributes in other Modules included in a Presentation State.
Table C.11.12-1. Presentation State Shutter Module Attributes
|
A single grayscale unsigned value used to replace those parts of the image occluded by the shutter, when rendered on a monochrome display. The units are specified in P-Values. Required if the Display Shutter Module or Bitmap Display Shutter Module is present. NoteThe requirement in this Module is type 1C, which overrides the type 3 in the Display Shutter Module. |
|||
|
A color triplet value used to replace those parts of the image occluded by the shutter, when rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1. Required if the Display Shutter Module or Bitmap Display Shutter Module is present and the SOP Class is other than Grayscale Softcopy Presentation State Storage. NoteThe requirement in this Module is type 1C, which overrides the type 3 in the Display Shutter Module and Bitmap Display Shutter Module. |
| DICOM PS3.3 2024e - Information Object Definitions |
|---|