| DICOM PS3.3 2024e - Information Object Definitions |
|---|
This Macro contains Attributes of a Waveform Filter Lookup Table.
Table C.10.13-1. Waveform Filter Lookup Table Macro Attributes
|
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
BCID 3044 “Waveform Filter Lookup Table Input Frequency Unit”. |
||
|
Encoding of filter response magnitude in the Lookup Table Data. |
|||
|
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
BCID 3045 “Waveform Filter Lookup Table Output Magnitude Unit”. |
||
|
See Section C.10.13.1 for further explanation. |
|||
Lookup Table Data consist of n triplets of double values: the first value is the input frequency. The second value is the magnitude. The third value is the phase shift of the signal resulting from filtering.
The input frequency values in the Lookup Table shall cover all frequencies from 0 to ½ of the sampling frequency. For input frequencies between two values in the Lookup Table an interpolation between the neighboring values can take place.
Frequency Encoding Code Sequence (003A,032C) defines the encoding of the first value, which could be:
Magnitude Encoding Code Sequence (003A,032D) defines the encoding of the second value, which could be:
Normalized magnitude is (ω - |H(ω)|)/ ωmax, which is the input signal amplitude minus the filter output signal amplitude both divided by the maximum input signal amplitude, and has a value between 0-1; (no units, because it is the ratio of input and output amplitudes).
| DICOM PS3.3 2024e - Information Object Definitions |
|---|