DICOM PS3.3 2024e - Information Object Definitions

A.51.3 Segmentation IOD Module Table

Table A.51-1 specifies the Modules of the Segmentation IOD.

Table A.51-1. Segmentation IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Segmentation Series

C.8.20.1

M

Clinical Trial Series

C.7.3.2

U

Frame of Reference

Frame of Reference

C.7.4.1

C - Required if Derivation Image Functional Group (C.7.6.16.2.6) is not present. May be present otherwise.

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

Acquisition

General Acquisition

C.7.10.1

U

Multi-Resolution Pyramid

Multi-Resolution Pyramid

C.7.11.1

U

Image

General Image

C.7.6.1

M

General Reference

C.12.4

U

Microscope Slide Layer Tile Organization

C.8.12.14

C - Required if Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise.

Image Pixel

C.7.6.3

M

Segmentation Image

C.8.20.2

M

Multi-frame Functional Groups

C.7.6.16

M

Multi-frame Dimension

C.7.6.17

M

Palette Color Lookup Table

C.7.9

C - Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR.

Specimen

C.7.6.22

U

Common Instance Reference

C.12.2

C - Required if Derivation Image Functional Group (C.7.6.16.2.6) is present.

ICC Profile

C.11.15

C - Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR.

SOP Common

C.12.1

M

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


Note

The Specimen Identification Module (Retired) was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS3.3-2008.

DICOM PS3.3 2024e - Information Object Definitions