| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table A.48-1 specifies the Modules of the Enhanced XRF Image IOD.
Table A.48-1. Enhanced XRF Image IOD Modules
The Overlay Plane Module, VOI LUT Module and Softcopy Presentation LUT Module shall not be used in a Standard Extended SOP Class of the Enhanced XRF Image.
The VOI LUT function is provided by a Frame VOI LUT Functional Group.
The Curve Module (Retired) was previously included in the list of Modules that shall not be present, but has been retired. It is still not permitted to be present. See PS3.3-2004.
The Specimen Identification Module (Retired) was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS3.3-2008.
Positioner Type (0018,1508) shall have the value COLUMN if the XA/XRF Acquisition Module is present.
| DICOM PS3.3 2024e - Information Object Definitions |
|---|