Table A.33.3-1 specifies the Modules of the Pseudo-color Softcopy Presentation State IOD.
Table A.33.3-1. Pseudo-Color Softcopy Presentation State IOD Modules
|
C - Required if the referenced image(s) are multi-frame and are to be subtracted |
|||
|
C - Required if a Display Shutter is to be applied to referenced image(s) and the Bitmap Display Shutter Module is not present |
|||
|
C - Required if a Display Shutter is to be applied to referenced image(s) and the Display Shutter Module is not present |
|||
|
C - Required if Overlay is to be applied to referenced image(s) or the Bitmap Display Shutter Module is present. |
|||
|
C - Required if referenced image contains overlay data that is to be displayed or Presentation State Instance contains Overlay data other than Bitmap Shutter |
|||
|
C - Required if Graphic Annotations are to be applied to referenced image(s) |
|||
|
C - Required if rotation or flipping are to be applied to referenced image(s) |
|||
|
C - Required if Graphic Annotations or Overlays are to be applied to referenced image(s) |
|||
|
C - Required if a Modality LUT is to be applied to referenced image(s) |
|||
|
C - Required if a VOI LUT is to be applied to referenced image(s) |
|||
In the Pseudo-color Softcopy Presentation State IOD, the Presentation Series Module specializes some Attributes of the General Series Module, the Presentation State Mask Module specializes some Attributes of the Mask Module, and the Presentation State Shutter Module specializes some Attributes of the Bitmap Display Shutter Module and Display Shutter Module.
The Softcopy Presentation LUT Module shall not be present in this IOD.