| DICOM PS3.3 2024e - Information Object Definitions |
|---|
Table A.27-1 specifies the Modules of the Digital Mammography X-Ray Image IOD.
Table A.27-1. Digital Mammography X-Ray Image IOD Modules
|
C - Required if multiple images are obtained without releasing breast compression |
|||
|
C - Required if graphic annotation is present - See Section A.27.3 |
|||
|
C - Required if Presentation Intent Type (0008,0068) is FOR PRESENTATION. Shall not be present otherwise. |
|||
The Curve Module (Retired) was previously included in this IOD but has been retired. See PS3.3-2004.
The Specimen Identification Module (Retired) was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS3.3-2008.
| DICOM PS3.3 2024e - Information Object Definitions |
|---|