N.6.2.6 Storage Commitment Service Configuration

Table N.6-7 lists Storage Commitment Service configuration parameters:

Table N.6-7. Storage Commitment Service Parameters

Local Configuration Parameters - Storage Commitment Service

Parameter

Configurable

Default Value

Comments

[This example shows the for a Storage Commitment SCU and SCP, e.g., a PACS.]

<<USER

SERVICE

FIXED>>

[If there is no default, leave blank.]

[Provide comments or Values/ranges if applicable.]

Calling AE Title (SCU)

SERVICE

STORE_AE

The system uses the same Calling AE Title as for the Storage SCU service by default

Called AE Title (SCP)

SERVICE

STORE_AE

The system uses the same Called AE Title as for the Storage SCP service by default

Port

SERVICE

104

TLS-Secured Port

FIXED

2762

Delay to send N-ACTION-RQ

SERVICE

300

Delay to send N-EVENT-REPORT-RQ

FIXED

immediately

As soon as the N-ACTION-RQ is received the system will initiate an Association to send the N-EVENT-REPORT

N-EVENT-REPORT on same Association

FIXED

asynchronous

When the system receives a N-ACTION, it will open a new Association to send the N-EVENT-REPORT.

When the system sends an N-ACTION, it expects to receive the N-EVENT-REPORT in a separate Association.

Remote Configuration Parameters - Storage Commitment Service

Parameter

Configurable

Default Value

Comments

<<USER

SERVICE

FIXED>>

[If there is no default, leave blank.]

[Provide comments or Values/ranges if applicable.]

Calling AE Title (SCU)

SERVICE

Called AE Title (SCP)

SERVICE

port

SERVICE

104

TLS-Secured Port

SERVICE

2762

Host

SERVICE


DICOM PS3.2 2024e - Conformance