DICOM PS3.2 2024e - Conformance

N.10.2 A.B.2 Echocardiography Procedure Result SR (TID 5200)

Table N.10-4 shows the encoding of content of a DICOM Echocardiography Procedure Report (TID 5200).

[Table N.10-4 shows how to document TID content usage, with TID 5200 as an example." Modify to match your product implementation, e.g., select supported concepts and Values, and add additional templates as needed.]

Table N.10-4. Echocardiography Procedure Report SR (TID 5200)

NL

Rel with Parent

VT

Concept Name

Source

Presence of Content Item

Values

TID

Comments

CONTAINER

EV (125200, DCM, "Adult Echocardiography Procedure Report")

5200

>

HAS CONCEPT MOD

CODE

(121049, DCM, "Language of Content Item and Descendants")

CONFIGURATION

(en, RFC3066, "English" )

1204

>>

HAS CONCEPT MOD

CODE

(121046, DCM, "Country of Language" )

CONFIGURATION

(US, ISO3166_1, "United States of America (the) " )

1204

>

HAS OBS CONTEXT

CODE

(121005, DCM, "Observer Type")

GENERATED

(121006, DCM, "Person")

1002

>

HAS OBS CONTEXT

PNAME

EV (121008, DCM, "Person Observer Name")

CONFIGURATION

1003

>

CONTAINS

CONTAINER

EV (121118, DCM, "Patient Characteristics"

GENERATED

5201

>>

CONTAINS

NUM

(121118, DCM, "Subject Age")

GENERATED

5201

Calculated from Date of Birth

>>

CONTAINS

CODE

EV (121032, DCM, "Subject Sex")

MWL

See CID 7455 "Sex"

5201

>>

CONTAINS

NUM

(8277-6, LN, "Body Surface Area")

GENERATED

5201

>>>

INFERED FROM

CODE

(8278-4, LN, "Body Surface Area Formula")

GENERATED

See CID 3663 "Body Surface Area Equations"

5201

>

CONTAINS

CONTAINER

(59776-5, LN, "Findings")

GENERATED

One Container for each supported Finding Site, see Section N.10.2.1

5202

The following rows are supported for all Finding Sites listed in Section N.10.2.1. Values for supported concepts are listed in the "Modifier" column of the Tables in the respective subsections of Section N.10.2.1.

>>

HAS CONCEPT MOD

CODE

(363698007, SCT "Finding Site"

GENERATED

See TID 5200 for supported Finding Sites

5202

>>

CONTAINS

CONTAINER

(125007, DCM, "Measurement Group)

5202

>>>

CONTAINS

NUM

See Section N.10.2.1 for measurements and supported Modifiers for each Finding Site

300

>>>>

HAS CONCEPT MOD

CODE

(370129005, SCT, "Measurement Method")

GENERATED

See CID 12227 "Echocardiography Measurement Method"

300

>>>>

HAS CONCEPT MOD

CODE

(363698007, SCT, "Finding Site")

GENERATED

See CID 12236 "Echo Anatomic Sites"

300

>>>>

HAS CONCEPT MOD

CODE

(26067400, SCT, "Flow Direction")

GENERATED

See CID 12221 "Flow Direction"

5203

>>>>

HAS CONCEPT MOD

CODE

(272517003, SCT, "Respiratory Cycle Point")

GENERATED

See CID 12234 "Respiration State"

5203

>>>>

HAS CONCEPT MOD

CODE

(272518008, SCT, "Cardiac Cycle Point")

GENERATED

See CID 12233 "Cardiac Phase"

5203

>>>>

HAS CONCEPT MOD

CODE

(399264008, SCT, "Image Mode")

GENERATED

See CID 12224 "Ultrasound Image Modes"

5203

>>>>

HAS CONCEPT MOD

CODE

(111031, DCM, "Image View")

GENERATED

See CID 12002 "Ultrasound Protocol Stage Types"

5203


N.10.2.1 A.B.2.1 Measurement Encoding

The following Sections provide a list of measurements encoded for each Finding Site.

[Since the lists of measurements can be fairly extensive, they can either be provided in a separate excel sheet minimally providing columns for

  • Label

  • The encoding of the measurement using Coding Scheme Designator, Code Value and Code Meaning

  • One column for each supported modifier (Image Mode, Image View, Measurement Method, Cardiac Cycle Point, …)

  • The unit code for the measurement using Coding Scheme Designator, Code Value and Code Meaning.]

[If you use an external document, state the following:]

Details about the supported measurements can be found at <link to external document>.

[If measurements are documented in this document, add for each supported Finding Site a subsection with all supported Measurements and their modifiers below following the examples shown.]

N.10.2.1.1 A.B.2.1.1 Left Ventricle

Table N.10-5 lists the measurements supported by <product>. The first column lists the label that is used on <products reporting screen>to select the respective measurements.

[Document all measurements supported on the product using the relevant measurements. Modify to match your product implementation, e.g., select supported concepts and Values, and add additional templates as needed. If private codes are used, indicate them through a 99_VENDOR_X Coding Scheme Designator, where VENDOR_X needs to be replaced with a vendor specific Value.]

[In the "Modifier" column list all supported modifiers by using the Concept Name Code from Table N.10-4 in Section N.10.2 and add a code for each Modifier Value.]

Table N.10-5. Left Ventricle Measurements

Label

Measurement

Modifier

Unit

Echo Section (TID 5202) - Left Ventricle, (363698007, SCT, "Finding Site") : (87878005, SCT, " Left Ventricle")

LV CI A2C MOD

(54993008, "SCT, Cardiac Index")

(399264008, SCT, "Image Mode")

(399064001, SCT, "2D mode")

(l/min/m2, UCUM, "l/min/m2")

(111031, DCM, "Image View")

(399232001, SCT, "Apical two chamber")

(370129005, SCT, "Measurement Method")

(125208, DCM, "Method of Disks, Single Plane")

LVID d PSAX A-P

(LVID_AP, 99VENDOR_X, Left Ventricle Internal Dimension A-P")

(272518008, SCT, "Cardiac Cycle Point")

(90892000, SCT, "Diastole")

(l/min/m2, UCUM, "l/min/m2")

(111031, DCM, "Image View")

(399271003, SCT, "Parasternal short axis at the Papillary Muscle level")

(399264008, SCT, "Image Mode")

(399064001, SCT, "2D mode")

…


N.10.2.1.2 A.B.2.1.2 Right Ventricle

Table N.10-6 list the measurements supported by <product>. The first column lists the label that is used on <products reporting screen>to select the respective measurements.

Table N.10-6. Right Ventricle Measurements

Label

Measurement

Modifier

Unit

Echo Section (TID 5202) - Right Ventricle, (363698007, SCT, "Finding Site") : (53085002, SCT, "Right Ventricle")

RV Area s A4C

(42798000, SCT "Area")

(272518008, SCT, "Cardiac Cycle Point")

(111973004, SCT, "Systole")

(cm2/m2, UCUM, "cm2/m2")

(111031, DCM, "Image View")

(399214001, SCT "Apical four chamber")

(399264008, SCT, "Image Mode")

(399064001, SCT, "2D mode")

(370129005, SCT, "Measurement Method")

(125208, DCM, "Method of Disks, Single Plane")

…


DICOM PS3.2 2024e - Conformance