DICOM PS3.3 2024d - Information Object Definitions |
---|
The following section contain Functional Group Macros specific to the Confocal Microscopy Image IOD and the Confocal Microscopy Tiled Pyramidal Image IOD.
Table C.8.35.4.1-1specifies the Attributes of the Confocal Microscopy Image Frame Type Macro
Table C.8.35.4.1-1. Confocal Microscopy Image Frame Type Macro Attributes
Identifies the characteristics of this Confocal Microscopy Image frame. |
|||
Type of Frame. A multi-valued Attribute analogous to Image Type (0008,0008). Enumerated Values and Defined Terms are the same as those for the four values of Image Type (0008,0008). See Section C.8.35.1.1.1. |
DICOM PS3.3 2024d - Information Object Definitions |
---|