DICOM PS3.3 2024d - Information Object Definitions

C.8.30 Corneal Topography Map Modules

C.8.30.1 Corneal Topography Map Series Module

Table C.8.30.1-1 specifies the Attributes of the Corneal Topography Map Series Module, which identify and describe general information about the Corneal Topography Map Series.

Table C.8.30.1-1. Corneal Topography Map Series Module Attributes

Attribute Name

Tag

Type

Attribute Description

Modality

(0008,0060)

1

Type of device, process or method that created the maps in this Series.

Enumerated Values:

OPM

See Section C.7.3.1.1.1 for further explanation.

Referenced Performed Procedure Step Sequence

(0008,1111)

1C

Uniquely identifies the Performed Procedure Step SOP Instance to which the Series is related (e.g., a Modality or General-Purpose Performed Procedure Step SOP Instance).

Only a single Item shall be included in this Sequence.

Required if the Modality Performed Procedure Step SOP Class, or General Purpose Performed Procedure Step SOP Class is supported.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Body Part Examined

(0018,0015)

1

Text description of the part of the body examined.

Enumerated Values:

EYE

Note

Some IODs support the Anatomic Region Sequence (0008,2218), which can provide a more comprehensive mechanism for specifying the body part being examined.


DICOM PS3.3 2024d - Information Object Definitions