DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.8.28.3-1 specifies the Attributes of the Ophthalmic Thickness Map Quality Rating Module, which describe the quality rating for the ophthalmic mapping.
Table C.8.28.3-1. Ophthalmic Thickness Map Quality Rating Module Attributes
Type of metric and metric value used to evaluate the quality of the ophthalmic mapping for grading and diagnostic purposes for this SOP Instance. |
|||
DCID 4243 “Ophthalmic Quality Metric Type” shall be used for Concept Name Code Sequence (0040,A043). |
|||
Quality threshold value and software algorithm used to provide the ophthalmic thickness map quality rating for this SOP Instance. |
|||
Quality rating threshold value for acceptable ophthalmic map. NoteThe units of this Attribute is the same as defined in Measurement Unit Code Sequence (0040,08EA) of the Ophthalmic Thickness Map Quality Rating Sequence (0022,1470). The threshold value is not the same as the Attribute Numeric Value (0049,A30A) of the Ophthalmic Thickness Map Quality Rating Sequence (0022,1470). Therefore, it conveys the least stringent value that is acceptable, not the actual rating for this SOP Instance. |
|||
>>Include Table 10-19 “Algorithm Identification Macro Attributes” |
DICOM PS3.3 2024d - Information Object Definitions |
---|