DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.8.25.12-1 defines the Attributes used to represent the subjective measurements of a patient's visual acuity. Usually each eye is measured individually, but sometimes only one eye is measured; sometimes the visual acuity of both eyes working together is measured. The visual acuity may be measured uncorrected, or with various refractive corrections, such as best corrected distance, pinhole, habitual near, etc
Table C.8.25.12-1. Visual Acuity Measurements Module Attributes
The viewing distance conditions under which a patient's vision was measured. |
|||
A Sequence that specifies the conditions under which a patient's vision was measured. |
|||
>Include Table 8.8-1 “Code Sequence Macro Attributes”. DCID 4216 “Ophthalmic Visual Acuity Type”. |
|||
The color of the background upon which optotypes are displayed for measurements of visual acuity. |
|||
The standardized test type (standard object of regard) used in measurements of visual acuity. |
|||
Required if the Optotype (0046,0094) Attribute value is LETTERS or NUMBERS or PICTURES. Provides additional information to define what optotype was used. |
|||
Specifies whether single or multiple optotypes were presented. |
|||
A Sequence that specifies the subjective measurement of a patient's visual acuity for the right eye. |
|||
>Include Table C.8.25.12-2 “Visual Acuity Measurements Macro Attributes” |
|||
A Sequence that specifies the subjective measurements of a patient's visual acuity for the left eye. |
|||
>Include Table C.8.25.12-2 “Visual Acuity Measurements Macro Attributes” |
|||
A Sequence that specifies a patient's subjective visual acuity measurements with both eyes open. |
|||
>Include Table C.8.25.12-2 “Visual Acuity Measurements Macro Attributes” |
Table C.8.25.12-2 specifies the Attributes of the Visual Acuity Measurements Macro.
Table C.8.25.12-2. Visual Acuity Measurements Macro Attributes
A patient's visual acuity specified in decimal. The value is derived from two values in a fraction where the numerator of the fraction is the nominal distance to the chart that the patient is reading. The denominator represents the line of smallest optotypes of which the patient can see more than half. Note
|
|||
The first value is a positive or negative integer numeric value such as +1, +2, +3, 0, -1, -2, or -3 used to indicate that the patient missed letters on the line referenced, or saw additional letters on the next smaller line. The second value is a positive or negative integer numeric value such as +1, +2, +3, 0, -1, -2, or -3, which in combination with the first value indicates that the patient both saw additional letters on the next smaller line and missed letters on the line referenced. Note
|
DICOM PS3.3 2024d - Information Object Definitions |
---|