DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.8.19.7-1 specifies the Attributes of the XA/XRF Multi-frame Presentation Module.
Table C.8.19.7-1. XA/XRF Multi-frame Presentation Module Attributes
Describes the preferred playback sequencing for a multi-frame image. |
|||
Sequence that specifies the display frame rate of a selected set of frames. The Items are ordered in increasing frame number. The range of the frames may not overlap and the ranges shall be adjacent. |
|||
The Frame Number of the first frame of the set of frames to be displayed in this Item. |
|||
The Frame Number of the last frame of the set of frames to be displayed in this Item. |
|||
A flag indicating that the range of frames in this Item may be skipped. |
|||
Recommended rate at which the frames of this Item should be displayed in frames/second. |
|||
Specifies the recommended viewing protocol(s). When this Attribute is present with a value, this value shall override the value of Recommended Viewing Mode (0028,1090) specified in the Mask Module. |
|||
Edge enhancement filter percentage that is recommended by the pixel data creator as filter presetting for display purposes. The value of 100 corresponds to the maximum filter strength that can be applied by a specific application displaying the image. |
|||
The percentage of visibility of the mask frame during a subtracted display. A value of 0 corresponds to subtracted display, a value of 100 corresponds to un-subtracted display (native). See Section C.8.19.7.1. Required if Recommended Viewing Mode (0028,1090) equals SUB. |
An example of the usage of the Mask Visibility Percentage (0028,9478) Attribute is illustrated below.
Assume that Psub is the output pixel level of a subtracted frame, its calculation can be expressed as followed:
Pcontrast: Pixel level of the contrast frame in the logarithmic domain
Pmask: Pixel level of the mask frame in the logarithmic domain
DICOM PS3.3 2024d - Information Object Definitions |
---|