DICOM PS3.3 2024d - Information Object Definitions

C.8.17.14 Ophthalmic Optical Coherence Tomography En Face Image Module

Table C.8.17.14-1 specifies the Attributes of the Ophthalmic Optical Coherence Tomography En Face Image Module, which describe the Ophthalmic Optical Coherence Tomography En Face Image Module.

Table

Table C.8.17.14-1. Ophthalmic Optical Coherence Tomography En Face Image Module Attributes

Attribute Name

Tag

Type

Attribute Description

Image Type

(0008,0008)

1

Image identification characteristics.

See Section C.8.17.14.1.5 for specialization.

Instance Number

(0020,0013)

1

A number that identifies this SOP Instance.

Bits Allocated

(0028,0100)

1

Number of bits allocated for each pixel sample. See Section C.8.17.14.1.6 for specialization.

Bits Stored

(0028,0101)

1

Number of bits stored for each pixel sample. See Section C.8.17.14.1.6 for specialization.

High Bit

(0028,0102)

1

Most significant bit for pixel sample data. Shall be one less than the value in Bits Stored (0028,0101). See Section C.8.17.14.1.6 for specialization.

Samples per Pixel

(0028,0002)

1

Number of samples (planes) in this image.

Enumerated Values:

1

Photometric Interpretation

(0028,0004)

1

Specifies the intended interpretation of the pixel data.

See Section C.8.17.14.1.4 for specialization.

Pixel Representation

(0028,0103)

1

Data representation of pixel samples.

Enumerated Values:

0000H

unsigned integer

Pixel Spacing

(0028,0030)

1

Nominal physical distance at the focal plane (in the retina) between the center of each pixel, specified by a numeric pair - adjacent row spacing (delimiter) adjacent column spacing in mm. See Section 10.7.1.3for further explanation of the value order.

Note

Since a patient's retina is curved and the image representation is planar, there can be an error in using Pixel Spacing (0028,0030) for measurements in the periphery of the image. En face imaging does not support wide field measurements.

Image Orientation (Patient)

(0020,0037)

1

The direction cosines of the first row and the first column with respect to the patient. See Section C.7.6.2.1.1 for further explanation.

Note

Since the coordinate system associated with the Frame of Reference is deformed and the selected en face data volume might be non-planar, the orientation will be a nominal real world orientation.

Ophthalmic Frame Location Sequence

(0022,0031)

1

Specifies the location of this image in terms of location on a referenced localizer image.

Only a single Item shall be included in this Sequence.

>Include Table 10-3 “Image SOP Instance Reference Macro Attributes”

>Reference Coordinates

(0022,0032)

1

Image coordinates for the points on the referenced image that correspond to the Top Left Hand Corner (TLHC) and the Bottom Right Hand Corner (BRHC) of the En Face Image.

Exactly two pairs of values where the first value of each pair is the row (vertical) offset and the second value of each pair is the column (horizontal) offset.

Specified with sub-pixel resolution such that the origin at the Top Left Hand Corner (TLHC) of the TLHC pixel is 0.0\0.0, the Bottom Right Hand Corner (BRHC) of the TLHC pixel is 1.0\1.0, and the BRHC of the BRHC pixel is Rows\Columns (see Figure C.10.5-1, except that row and column order is reversed). The values must be within the range 0\0 to Rows\Columns of the referenced image.

See Section C.8.17.10.1.1. Alignment of an En Face Image is equivalent to that of a transverse OPT Image.

Content Time

(0008,0033)

1

The time the image pixel data creation started.

Content Date

(0008,0023)

1

The date the image pixel data creation started.

Source Image Sequence

(0008,2112)

1

A Sequence that identifies the Images that were used to derive this Image.

One or more Items shall be included in this Sequence.

See Section C.12.4.1.2 and Section C.8.17.14.1.1 for further explanation.

>Include Table 10-3 “Image SOP Instance Reference Macro Attributes”

>Purpose of Reference Code Sequence

(0040,A170)

1

Describes the purpose for which the reference is made, that is what role the source image or frame(s) played in the derivation of this image.

Only a single Item shall be included in this Sequence.

>>Include Table 8.8-1 “Code Sequence Macro Attributes”

DCID 7202 “Source Image Purpose of Reference”.

If the derived en face image is based upon OPT structure information, the Concept Code shall be (128250, DCM, "Structural image for image processing").

If the derived en face image is based upon OPT flow information, the Concept Code shall be (128251, DCM, "Flow image for image processing").

Derivation Algorithm Sequence

(0022,1612)

1

Software algorithm that performed the derivation.

Only a single Item shall be included in this Sequence.

>Include Table 10-19 “Algorithm Identification Macro Attributes”

DCID 4274 “En Face Processing Algorithm Family” shall be used for Algorithm Family Code Sequence (0066,002F)

Note

Additional processing, such as artifact removal, that are used in the derivation but not strictly part of the algorithm, can be described in Algorithm Parameters (0066,0032).

Ophthalmic Image Type Code Sequence

(0022,1615)

1

En face image type used to identify this SOP Instance

Only a single Item shall be included in this Sequence.

See Section C.8.17.14.1.3 for further explanation.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

DCID 4271 “En Face Image Type”

Ophthalmic Image Type Description

(0022,1616)

3

Description of the en face image type.

Window Center

(0028,1050)

1

Window Center for display.

See Section C.11.2.1.2 for further explanation.

Note

When Bits Allocated (0028,0100) is 8, the value is typically always set to 127 or 128.

Window Width

(0028,1051)

1

Window Width for display. See Section C.11.2.1.2 for further explanation.

Note

When Bits Allocated (0028,0100) is 8, the value is typically always set to 256.

Ophthalmic FOV

(0022,1517)

3

The horizontal field of view used to capture the ophthalmic image, in degrees. The field of view is the maximum image size displayed on the image plane, expressed as the angle subtended at the exit pupil of the eye by the maximum dimension 2r (where r equals the radius).

Ophthalmic En Face Volume Descriptor Sequence

(0022,1627)

1

Description of the volume or boundary surfaces used to select the en face image data from the source image(s).

One or two Items shall be included in this Sequence.

See Section C.8.17.14.1.2 for further explanation.

>Ophthalmic En Face Volume Descriptor Scope

(0022,1629)

1

Part of the en face volume described by this Item.

Enumerated Values:

ANTERIOR

Anterior surface of volume

POSTERIOR

Posterior surface of volume

ENTIRE

Entire volume

If value is ENTIRE, this Item shall be the only Item in the Ophthalmic En Face Volume Descriptor Sequence (0022,1627). Otherwise, two Items shall be included in the Ophthalmic En Face Volume Descriptor Sequence (0022,1627), one with value ANTERIOR and the other with value POSTERIOR.

Note

Anterior and posterior surfaces are defined in terms of the orientation of the en face volume within the patient coordinate system in the source image(s).

>Referenced Segmentation Sequence

(0008,114C)

1C

Reference to the segmentations used in the selection of the en face data.

One or more Items shall be included in this Sequence.

Required if segmentation is used to select the en face data volume or surface.

See Section C.8.17.14.1.2 for further explanation.

>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes”

Reference to a segmentation SOP Instance and one or more segments thereof in Referenced Segment Number (0062,000B). Referenced Segment Number (0062,000B) shall be present, even if the referenced segmentation SOP Instance contains only a single segment.

Note

The SOP Class of the segmentation is not constrained.

>>Segmented Property Type Code Sequence

(0062,000F)

1

Sequence defining the specific property the segmentation represents. The Items in this Sequence shall be copied from the Segmented Property Type Code Sequence (0062,000F) of the referenced segmentation.

The number of Items in this Sequence shall equal the number of values in Referenced Segment Number (0062,000B).

>>>Include Table 8.8-1 “Code Sequence Macro Attributes”

>>>Segmented Property Type Modifier Code Sequence

(0062,0011)

3

Sequence defining the modifier of the property type of this segment.

One or more Items are permitted in this Sequence

>>>>Include Table 8.8-1 “Code Sequence Macro Attributes”

>Surface Offset

(0066,0005)

1

Offset in pixels from the referenced segmentation surface, in the direction from the top towards the bottom of the source image frames. If no referenced segmentation surface is specified in this Item of Ophthalmic En Face Volume Descriptor Sequence (0022,1627), the offset is from the top of the source image frames.

If no offset is used the value is set to 0.

See Section C.8.17.14.1.2.

>Surface Processing Description

(0066,000B)

3

A description of processing performed to construct the surface, such as interpolation between referenced segmented surfaces.

Ophthalmic Axial Length

(0022,1019)

3

The axial length measurement, in mm.

Ophthalmic Axial Length Method

(0022,1515)

3

The method used to obtain the Ophthalmic Axial Length.

Enumerated values:

MEASURED

Measured axial length.

ESTIMATED

An estimated value based upon performing the examination (i.e., based upon surrogate markers of axial length).

POPULATION

A length that represents a population norm (i.e., not based upon a measured axial length or surrogate markers of axial length).

Lossy Image Compression

(0028,2110)

1

Specifies whether an Image has undergone lossy compression (at a point in its lifetime).

Enumerated Values:

00

Image has not been subjected to lossy compression.

01

Image has been subjected to lossy compression.

Once this value has been set to "01" it shall not be reset.

See Section C.7.6.1.1.5.

Lossy Image Compression Ratio

(0028,2112)

1C

Describes the approximate lossy compression ratio(s) that have been applied to this image.

See Section C.7.6.1.1.5.2.

Required if Lossy Image Compression (0028,2110) is "01".

Lossy Image Compression Method

(0028,2114)

1C

A label for the lossy compression method(s) that have been applied to this image.

See Section C.7.6.1.1.5.1.

Required if Lossy Image Compression (0028,2110) is "01".

Presentation LUT Shape

(2050,0020)

1C

Specifies an identity transformation for the Presentation LUT, such that the output of all grayscale transformations defined in the IOD containing this Module are defined to be P-Values.

Enumerated Values:

IDENTITY

output is in P-Values

Required if Photometric Interpretation (0028,0004) is MONOCHROME2.

Calibration Image

(0050,0004)

3

Indicates whether a reference object (phantom) of known size is present in the image and was used for calibration.

Enumerated Values:

YES

NO

Burned In Annotation

(0028,0301)

1

Indicates whether or not image contains sufficient burned in annotation to identify the patient and date the image was acquired.

Enumerated Values:

YES

NO

Recognizable Visual Features

(0028,0302)

1

Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient.

Enumerated Values:

YES

NO


C.8.17.14.1 Ophthalmic Optical Coherence Tomography En Face Image Module Attribute Descriptions

In this Section, the term "surface segmentation" (uncapitalized) is a generic reference to any type of segmentation that describes a surface. It includes both the Surface Segmentation IOD or SOP Class (capitalized) and the Height Map Segmentation IOD or SOP Class.

C.8.17.14.1.1 Source Image Sequence

An OCT en face image is derived from images obtained using OCT technology. The Source Image Sequence (0008,2112) shall convey the SOP Instances used to derive this en face SOP Instance.

If Attribute Purpose of Reference Code Sequence (0040,A170) is set to (128250, DCM, "Structural image for image processing"), the Source Image Sequence will reference an Ophthalmic Tomography SOP Instance.

If Attribute Purpose of Reference Code Sequence (0040,A170) is set to (128251, DCM, "Flow image for image processing"), the Source Image Sequence will reference an Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instance.

A typical example of the image processing stages performed to generate en face images is shown in Figure C.8.17.14-1.

Example of the Image Process Performed to Generate En Face Images

Figure C.8.17.14-1. Example of the Image Process Performed to Generate En Face Images


Figure Legend:

  1. OCT proprietary B-scan data (possibly a DICOM Raw Data Instance)

  2. Volumetric structural ophthalmic tomography image (Ophthalmic Tomography Image Instance)

  3. OCT angiographic flow volume information (Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Instance)

  4. OCT surface segmentation (e.g., Height Map Segmentation or Surface Segmentation Instance)

  5. Structural en face image (Ophthalmic Optical Coherence Tomography En Face Image Instance)

  6. En face angiographic flow image (Ophthalmic Optical Coherence Tomography En Face Image Instance)

Stage 1: OCT technology is used to acquire a volumetric data set from a retinal region of interest. This volumetric data set (A) consists of multiple B-scans in a raster pattern, and multiple frames are acquired at each B-scan location. The B-scans are acquired in the manufacturer's proprietary format for analysis and storage. If this information is stored in DICOM, it can use the Raw Data Storage SOP Class.

Stage 2: The OCT proprietary B-scan data (A) (or DICOM Raw Data SOP Instance) is then analyzed to derive the volumetric structural ophthalmic tomography image (B). From (B) one or more OCT surface segmentations (D) are generated to delineate the anatomical boundaries. The difference in signal between the frames of each individual B-scan is analyzed to produce the OCT angiographic flow volume information (C).

Stage 3: Clinicians typically make their assessment based upon two types of OCT en face images. The structural OCT en face image (E) is derived by using pixel information in (B) and two surface segmentationes (D). The OCT angiographic flow en face image (F) may be derived using the OCT angiographic flow volume information (C) and the OCT surface segmentation (D).

En face images are typically derived by the acquisition modality that generated the Ophthalmic Tomography Image, Surface Segmentation and Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instances or by image workstations that received the respective Ophthalmic Image, Surface Segmentation and Ophthalmic Optical Tomography B-scan Volume Analysis SOP Instances via DICOM Storage.

Note

Image workstations receiving ophthalmic tomography images may choose to evaluate the structural ophthalmic tomography image and generate a different set of segmented surfaces than defined by an acquisition device. The surface segmentation information can be stored in a separate Segmentation SOP Instance.

The Ophthalmic Tomography Image, Surface Segmentation, Ophthalmic Optical Coherence Tomography B-scan Volume Analysis and the Ophthalmic Optical Coherence Tomography En Face Image SOP Instances all reside in different DICOM Series. They share the same spatial Frame of Reference which is identified in Attribute Frame of Reference UID (0020,0052) (i.e., the value of Frame of Reference UID (0020,0052) is the same in each SOP Instance). Figure C.8.17.14-2 illustrates the relationships between the OCT angiography based SOP Instances.

Relationships Between OCT-A Based SOP Instances

Figure C.8.17.14-2. Relationships Between OCT-A Based SOP Instances


C.8.17.14.1.2 Ophthalmic En Face Volume Descriptor Sequence

The Items of the Ophthalmic En Face Volume Descriptor Sequence (0022,1627) identify the parameters used to select the data volume (slab) from the source image(s) for the derived en face image. The en face image data may be selected by a volumetric segmentation specified in a single Item of the Ophthalmic En Face Volume Descriptor Sequence (0022,1627), or by specifying an anterior and a posterior surface respectively in two Items.

Anterior and posterior surfaces may each be specified by

  • a referenced surface segmentation,

  • a combination (e.g., interpolation) of two referenced surface segments,

  • a fixed offset from a referenced surface segmentation, or

  • a fixed offset from the top of the source image frames.

If a referenced segmentation does not extend to the full pixel matrix of the en face image, the en face pixel values outside the extent of the segment are implementation specific. If the surfaces intersect, or reverse (anterior surface behind the posterior), the en face pixel values at the points of intersection or reversal are implementation specific.

Note

Such pixels may be represented with values in the Pixel Padding range.

Surface Offset (0066,0005) specifies an offset from the referenced segmentation surface to the surface of the en face image slab. If no segmentation is referenced in the same Item of the Ophthalmic En Face Volume Descriptor Sequence (0022,1627), the offset is from the top of the source image frames (i.e., the boundary surface is flat relative to the source image volume). The offset is a fractional number of pixels relative to the top of the frames of the source image, i.e., a positive number indicates offset toward the bottom of the frame.

Note

  1. The two en face data boundary surfaces may be specified relative to the same referenced segmentation surface, potentially with different offsets. The reference is duplicated in the two Items of the Ophthalmic En Face Volume Descriptor Sequence (0022,1627).

  2. An ENTIRE en face data volume may be specified by two referenced surface segments in a single Item of the Ophthalmic En Face Volume Descriptor Sequence (0022,1627). The two referenced segments may be specified in two Items of the Referenced Segmentation Sequence (0008,114C), or by a single Item that identifies two surfaces in Referenced Segment Number (0062,000B).

  3. An en face data boundary surface may be specified by a combination (e.g., interpolation) of multiple referenced segments. The multiple referenced segments may be specified in multiple Items of the Referenced Segmentation Sequence (0008,114C), or by a single Item that identifies multiple surfaces in Referenced Segment Number (0062,000B). The nature of the combination may be described in Surface Processing Description (0066,000B).

  4. An application that wishes to specify an offset that has been determined in real world units, e.g., 0.015 mm above the segmented surface, must calculate the offset in pixels by using the pixel measures (row spacing) of the segmentation derivation image to convert from real world distances to fractional number of pixels, and use that value in the Surface Offset (0066,0005) Attribute.

  5. This Module allows the creating application to record its processing for purposes of provenance and traceability. It does not necessarily provide sufficient information for a receiving application to reproduce an identical en face image.

C.8.17.14.1.3 Ophthalmic Image Type Code Sequence

Implementations may generate many different types of derived en face images. Figure C.8.17.14-3 illustrates various derived enface image types. The Ophthalmic Image Type Code Sequence (0022,1615) is used to identify the type of derived en face image.

Examples of En Face Images Types

Figure C.8.17.14-3. Examples of En Face Images Types


C.8.17.14.1.4 Photometric Interpretation

Specifies the intended interpretation of the pixel data.

Enumerated Values:

MONOCHROME2

PALETTE COLOR

C.8.17.14.1.5 Image Type

The Image Type (0008,0008) Attribute (General Image Module, Section C.7.6.1) identifies important image characteristics in a multiple valued Data Element. For the Ophthalmic Optical Tomography En Face Image IOD, Image Type is specified as a Type 1 Attribute and further specialized as follows:

  • Value 1 shall identify the Pixel Data Characteristics in accordance with Section C.7.6.1.1.2.

    Enumerated Values for Value 1:

    DERIVED

  • Value 2 shall identify the Patient Examination Characteristics in accordance with Section C.7.6.1.1.2.

    Enumerated Values for Value 2:

    PRIMARY

  • Value 3 shall identify any Image IOD specific specialization in accordance with Section C.7.6.1.1.2 (optional).

    Defined Terms for Value 3:

    MONTAGE

    identifies a Montage Image

  • Other Values that are implementation specific in accordance with Section C.7.6.1.1.2 (optional)

Note

A Montage Image is constructed out of several individual images, which also can be exchanged separately. The images used to create the montage image will be included in the Source Image Sequence (0008,2112) if those images are also exchanged. A Montage Image is identified as Image Type DERIVED\PRIMARY\MONTAGE.

C.8.17.14.1.6 Bits Allocated, Bits Stored, and High Bit

These Attributes shall be determined based upon the Photometric Interpretation (0028,0004):

Photometric Interpretation (0028,0004)

Bits Allocated (0028,0100)

Bits Stored (0028,0101)

High Bit (0028,0102)

MONOCHROME2

PALETTE COLOR

8

8

7

16

12

11

16

16

15

C.8.17.14.1.7 Relationship Between Ophthalmic Tomography Image and Ophthalmic Optical Coherence Tomography B-scan Volume Analysis IODs

When generating an angiographic en face Image SOP Instance implementations need to understand the relationship between the Ophthalmic Tomography Image SOP Instance(s) and the Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instance.

The Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instance, which is a multi-frame SOP Instance, references one or more Ophthalmic Tomography Image SOP Instances using the Derivation Image Macro. The Derivation Image Macro defines Attributes at the Frame Level (i.e. each frame in the Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instance, references an Ophthalmic Tomography Image SOP Instance and the Ophthalmic Tomography Image SOP Instance frame number that was used to generate the specific Ophthalmic Optical Coherence Tomography B-scan Volume Analysis frame).

Examples are shown in Section UUU.2 “Relationship Between Ophthalmic Tomography Image and Ophthalmic Optical Coherence Tomography B-scan Volume Analysis IODs” in PS3.17.

DICOM PS3.3 2024d - Information Object Definitions