DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.8-100c specifies the Attributes of the Functional MR Macro.
Table C.8-100c. Functional MR Macro Attributes
Functional Studies require synchronization between the Acquisition Modality and external devices to coordinate paradigm presentation by an external system with image acquisition. It is required that the Acquisition Time Synchronized (0018,1800) shall be "Y".
The Functional Sync Pulse (0018,9623) is the date-time stamp of the pulse at the start of the acquired volume(s) in the SOP Instance. All frames with the same combination of Stack ID (0020,9056) and Temporal Position Index (0020,9128) share the same time stamp.
DICOM PS3.3 2024d - Information Object Definitions |
---|