DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.8-87 specifies the Attributes of the MR Pulse Sequence Module.
The primary purpose of this Module is to identify the pulse sequence and variations on that, which was used in creation of the image. Terminology is intended to be neutral, and allow equivalent sequences provided by different vendors to be classified the same.
Table C.8-87. MR Pulse Sequence Module Attributes
Name of the pulse sequence for annotation purposes. Potentially vendor-specific name. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Identification of spatial data encoding scheme. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Echo category of pulse sequences. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Multiple Spin Echo category of pulse sequence used to collect different lines in k-space for a single frame. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED and Echo Pulse Sequence (0018,9008) equals SPIN or BOTH. Otherwise may be present if Image Type (0008,0008) Value 1 is DERIVED and Echo Pulse Sequence (0018,9008) equals SPIN or BOTH. |
|||
Technique that simultaneously excites several volumes. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Phase contrast pulse sequence is a pulse sequence in which the flowing spins are velocity encoded in phase. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Velocity encoding directions used for acquisition. |
|||
The direction cosines of the velocity encoding vector with respect to the patient. See Section C.7.6.2.1.1 for further explanation. |
|||
Time of Flight contrast is created by the inflow of blood in the saturated plane. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Arterial Spin Labeling contrast technique. Enumerated Values: Required if Image Type (0008,0008) Value 3 is ASL. May be present otherwise. |
|||
Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Echo Planar category of Pulse Sequences. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Saturation recovery pulse sequence. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Spectrally Selected Suppression. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Geometry category of k-Space traversal. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Rectilinear phase encode reordering. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED and Geometry of k-Space Traversal (0018,9032) equals RECTILINEAR. Otherwise may be present if Image Type (0008,0008) Value 1 is DERIVED and Geometry of k-Space Traversal (0018,9032) equals RECTILINEAR. |
|||
Segmented k-Space traversal. If Geometry of k-Space Traversal is rectilinear, multiple lines can be acquired at one time. If Geometry of k-Space Traversal is spiral or radial, paths can be interleaved and acquired at one time. Enumerated Values: Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|||
Coverage of k-Space in the ky-kz plane. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED and MR Acquisition Type (0018,0023) equals 3D. Otherwise may be present if Image Type (0008,0008) Value 1 is DERIVED and MR Acquisition Type (0018,0023) equals 3D. |
|||
Number of interleaves or shots. Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
DICOM PS3.3 2024d - Information Object Definitions |
---|