DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.8-72 specifies the Attributes of the DX Positioning Module, which describe the positioning used in acquiring Digital X-Ray Images.
Table C.8-72. DX Positioning Module Attributes
A Sequence that describes the radiographic method of patient, tube and detector positioning to achieve a well described projection or view. Only a single Item is permitted in this Sequence. Shall be consistent with the other Attributes in this Module, if present, but may more specifically describe the image acquisition. |
|||
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
|||
Description of imaging subject's position relative to the equipment. See Section C.7.3.1.1.2 for Defined Terms and further explanation. If present, shall be consistent with Patient Gantry Relationship Code Sequence (0054,0414) and Patient Orientation Modifier Code Sequence (0054,0412). |
|||
Radiographic view of the image relative to the imaging subject's orientation. Shall be consistent with View Code Sequence (0054,0220). See Section C.8.11.5.1.1 for further explanation. |
|||
Sequence that describes the projection of the anatomic region of interest on the image receptor. NoteIt is strongly recommended that this Attribute be present, in order to ensure that images may be positioned correctly relative to one another for display. Shall be consistent with View Position (0018,5101). See Section C.8.11.5.1.1 for further explanation. |
|||
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
|||
Sequence that describes the orientation of the patient with respect to gravity. See Section C.8.11.5.1.2 for further explanation. |
|||
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
|||
Required if needed to fully specify the orientation of the patient with respect to gravity. |
|||
Sequence that describes the orientation of the patient with respect to the gantry. |
|||
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
|||
Distance in mm from source to the table, support or bucky side that is closest to the Imaging Subject, as measured along the central ray of the X-Ray beam. Note
See Section C.8.11.7 for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC. |
|||
Distance in mm from source to detector center. See Section C.8.11.7 for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC. |
|||
Ratio of Source Image Receptor Distance (SID) over Source Object Distance (SOD). |
|||
Position of the X-Ray beam about the patient from the RAO to LAO direction where movement from RAO to vertical is positive, if Positioner Type (0018,1508) is CARM. See Section C.8.7.5 for further explanation if Positioner Type (0018,1508) is CARM. See Section C.8.11.7 for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC. |
|||
Position of the X-Ray beam about the patient from the CAU to CRA direction where movement from CAU to vertical is positive, if Positioner Type (0018,1508) is CARM. See Section C.8.7.5 for further explanation if Positioner Type (0018,1508) is CARM. See Section C.8.11.7 for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC. |
|||
Angle of the X-Ray beam in the row direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted toward higher numbered columns. Negative values indicate that the X-Ray beam is tilted toward lower numbered columns. See Section C.8.7.5 for further explanation. See Section C.8.11.7 for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC. |
|||
Angle of the X-Ray beam in the column direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted toward lower numbered rows. Negative values indicate that the X-Ray beam is tilted toward higher numbered rows. See Section C.8.7.5 for further explanation. See Section C.8.11.7 for explanation if Positioner Type (0018,1508) is MAMMOGRAPHIC. |
|||
Angle of the X-Ray beam in degree relative to an orthogonal axis to the detector plane. Positive values indicate that the tilt is toward the head of the table. |
|||
Angle of table plane in degrees relative to horizontal plane [Gravity plane]. Positive values indicate that the head of the table is upward. |
|||
The average thickness in mm of the body part examined when compressed, if compression has been applied during exposure. |
|||
The compression force applied to the body part during exposure, measured in Newtons. |
|||
The average compression pressure applied to the body part during exposure, calculated in kPa. |
|||
The area of the body part to which compression has been applied during exposure, measured in mm2. |
|||
Description of the compression paddle, if compression was applied to the body part during exposure. |
View Code Sequence (0054,0220) replaces the function of View Position (0018,5101), and describes the radiographic view of the image relative to the real-world patient orientation as described in Annex A “Explanation of Patient Orientation (Normative)” in PS3.17.
DICOM PS3.3 2024d - Information Object Definitions |
---|