DICOM PS3.3 2024d - Information Object Definitions |
---|
The RT Equipment Mapping and Plan Reference Macro allows to specify the geometric relationship of the imaging acquisition device and/or patient position to an RT Treatment Delivery Device.
Table C.36.2.4.12-1. RT Equipment Mapping and Plan Reference Macro Attributes
Frame of Reference identifier for the Treatment Delivery Device. Required if either Imaging Equipment to Treatment Delivery Device Relationship Sequence (300A,07A1) or Patient to Equipment Relationship Sequence (300A,07A0) is present. See Section C.36.12.1 and Section C.36.2.4.12.1.1. |
|||
Imaging Equipment to Treatment Delivery Device Relationship Sequence |
The relationship between the coordinate system of the Imaging Device and coordinate system of the Treatment Delivery Device. Only a single Item is permitted in this Sequence. See Section 10.39.1.1. |
||
A rigid, homogeneous 4x4 transformation matrix that maps the coordinate system of the imaging equipment to the coordinate system of the RT treatment device as identified by Equipment Frame of Reference UID (300A,0675). Matrix elements shall be listed in row-major order. |
|||
Translational and rotational parameters for the Imaging Device. Shall be consistent with the Image to Equipment Mapping Matrix (0028,9520). |
|||
Isocenter coordinates (x,y,z), in mm. Specifies the location of the machine isocenter in the Patient-Based Coordinate System associated with the Frame of Reference. It allows transformation from the Equipment-Based Coordinate System to the Patient-Based Coordinate System. |
|||
The relationship between the Patient Frame of Reference and the coordinate system of the Treatment Delivery Device. |
|||
A rigid, homogeneous 4x4 transformation matrix that maps the patient coordinate space in the Frame of Reference used for the patient model to the coordinate system of the RT Treatment Delivery Device as identified by Equipment Frame of Reference UID (300A,0675). Matrix elements shall be listed in row-major order. See Section 10.39.1.1, Section 10.39.1.2 and Section C.7.6.21.1. |
|||
Translational and rotational parameters for the Patient Support device. Shall be consistent with the Image to Equipment Mapping Matrix (0028,9520). |
|||
Procedures and devices required to prepare a patient for treatment. Only a single Item is permitted in this Sequence. See Section C.8.8.12.1.3. |
|||
>Include Table C.36.2.3.4-1 “RT Patient Treatment Preparation Macro Attributes”. |
|||
RT Plan Instance associated with the acquisition of this Instance. |
|||
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
Beams in referenced RT Plan Instance that are relevant for the acquisition of this Instance. |
|||
Uniquely identifies Beam specified by Beam Number (300A,00C0) in Beam Sequence (300A,00B0) within referenced RT Plan or in Ion Beam Sequence (300A,03A2) within RT Ion Plan. |
The Equipment Frame of Reference UID (300A,0675) identifies the Equipment Coordinate System for a Treatment Delivery Device.
If the Equipment Frame of Reference UID of a Treatment Delivery Device is not a Well-known Frame of Reference defined in Section C.36.12.2, it is expected that the coordinate systems and the axis parameters of the Treatment Delivery Device are documented in the Conformance Statement for that device.
DICOM PS3.3 2024d - Information Object Definitions |
---|