DICOM PS3.3 2024d - Information Object Definitions
C.36.2.4.10 RT Image Frame Radiation Acquisition Parameters Macro

The Radiation Image Acquisition Parameters Macro contains parameters specifying details of generation of the radiation used for image acquisition.

Table C.36.2.4.10-1. RT Image Frame Radiation Acquisition Parameters Macro Attributes

Attribute Name

Tag

Type

Attribute Description

RT Image Frame Radiation Acquisition Sequence

(3002,010C)

1

The acquisition parameters describing the generation of radiation of the frame.

Only a single Item shall be included in this Sequence.

>RT Image Frame kV Radiation Acquisition Sequence

(3002,010A)

1C

The kV photon acquisition parameters of the frame.

Required if RT Image Frame MV Radiation Acquisition Sequence (3002,010B) is not present.

Only a single Item shall be included in this Sequence.

>>Include Table C.36.2.4.6-1 “KV Radiation Image Acquisition Parameters Macro Attributes”

>RT Image Frame MV Radiation Acquisition Sequence

(3002,010B)

1C

The MV photon or particle acquisition parameters of the frame.

Required if RT Image Frame kV Radiation Acquisition Sequence (3002,010A) is not present.

Only a single Item shall be included in this Sequence.

>>Include Table C.36.2.4.7-1 “MV Radiation Image Acquisition Parameters Macro Attributes”


DICOM PS3.3 2024d - Information Object Definitions