DICOM PS3.3 2024d - Information Object Definitions
C.36.2.3.4 RT Patient Treatment Preparation Macro

Table C.36.2.3.4-1. RT Patient Treatment Preparation Macro Attributes

Attribute Name

Tag

Type

Description

Patient Treatment Preparation Method Code Sequence

(300A,078D)

1

Method used to prepare the Patient for treatment.

Only a single Item shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”.

BCID 9571 “Patient Treatment Preparation Method”.

Patient Treatment Preparation Method Description

(300A,0792)

3

User-defined description of patient treatment preparation technique.

Referenced Patient Setup Photo Sequence

(300A,078C)

3

References to photos illustrating the patient setup.

One or more Items are permitted in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

>Patient Setup Photo Description

(300A,0794)

2

User-defined description of patient setup photo.

>Referenced Patient Setup Procedure Index

(300A,0796)

1C

The value of Patient Treatment Preparation Procedure Index (300A,0795) from Patient Treatment Preparation Procedure Sequence (300A,0790) corresponding to the Patient Treatment Preparation Procedure to which this Sequence Item refers.

Required if this Patient Setup Photo is associated with a Patient Treatment Preparation Procedure.

Patient Treatment Preparation Procedure Sequence

(300A,0790)

2

Procedures for preparing for the treatment of the patient.

Zero or more Items shall be included in this Sequence.

>Patient Treatment Preparation Procedure Index

(300A,0795)

1

Index of the Patient Treatment Preparation Procedure in this Sequence.

The value shall start at 1 and increase monotonically by 1.

>Patient Treatment Preparation Procedure Code Sequence

(300A,0791)

1

Procedure performed to prepare the patient for treatment.

Only a single Item shall be present in this Sequence.

>>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 9577 “Patient Treatment Preparation Procedure”.

>Patient Treatment Preparation Device Sequence

(300A,078F)

3

Devices used for preparing the patient for treatment.

Only a single Item is permitted in this Sequence.

>>Include Table C.36.2.2.3-1 “RT Accessory Device Identification Macro Attributes”

BCID 9573 “Patient Treatment Preparation Device”.

>Patient Treatment Preparation Procedure Parameter Description

(300A,078E)

2

User-defined description of patient treatment preparation procedure parameter(s).

>Patient Treatment Preparation Procedure Parameter Sequence

(300A,0793)

2

Parameters for the Patient Treatment Preparation Procedure.

Zero or more Items shall be included in this Sequence.

>>Include Table 10.2.1-1 “Content Item with Modifiers Macro Attributes”

See C.36.2.3.4.1.1 for Baseline TID.


C.36.2.3.4.1 RT Patient Treatment Preparation Macro Attribute Description
C.36.2.3.4.1.1 Patient Treatment Preparation Procedure

In Table C.36.2.3.4.1.1-1, if the Patient Treatment Preparation Procedure Code Sequence (300A,0791) has the code value specified in the left column, the Template for Patient Treatment Preparation Procedure Parameter Sequence (300A,0793) shall be the one specified in the right column.

Table C.36.2.3.4.1.1-1. Patient Treatment Preparation Procedure Parameter Codes


DICOM PS3.3 2024d - Information Object Definitions