DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.36.2.2.11-1 specifies the Attributes of the Wedge Positions Macro, which define the positions of Wedges used in a specific Control Point or set of Control Points.
Table C.36.2.2.11-1. Wedge Positions Macro Attributes
Number of Wedge Positions defined in the Wedge Position Sequence (300A,0116). Required if Number of Wedges (300A,00D0) is present and has a non-zero value. |
|||
Position for each Wedge for the current Control Point. Required if Number of Wedge Positions (300A,0655) is present and has a non-zero value and the conditions in Section C.36.2.2.5.1.1 are satisfied. The number of Items included in this Sequence shall equal the value of Number of Wedge Positions (300A,0655). |
|||
The value of Device Index (3010,0039) in Wedge Definition Sequence (300A,0651) for the Wedge being used. |
|||
Closest distance in mm from the central axis of the beam along the wedge angle direction to the thin edge as projected on the Beam Modifier Definition Plane defined by the RT Beam Modifier Definition Distance (300A,0688). The value is negative if the thin edge is located on the positive y-axis, and is positive otherwise. |
DICOM PS3.3 2024d - Information Object Definitions |
---|