DICOM PS3.3 2024d - Information Object Definitions

C.2 Patient Modules

The following Sections specify Modules used for Patient management.

C.2.1 Patient Relationship Module

Table C.2-1 specifies the Attributes of the Patient Relationship Module, which reference SOP Instances related to this SOP Class.

Table C.2-1. Patient Relationship Module Attributes

Attribute Name

Tag

Attribute Description

Referenced Study Sequence

(0008,1110)

Uniquely identifies the Studies associated with the Patient. One or more Items shall be included in this Sequence.

See Section 10.6.1.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Referenced Visit Sequence

(0008,1125)

Uniquely identifies the Visits associated with this Patient. One or more Items shall be included in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”


DICOM PS3.3 2024d - Information Object Definitions