DICOM PS3.3 2024d - Information Object Definitions |
---|
Table C.13-3. Basic Film Box Presentation Module Attributes
Enumerated Values:
|
||
Identification of annotation display format. The definition of the annotation display formats and the annotation box position Sequence are defined in the Conformance Statement. |
||
Defined Terms: |
||
Interpolation type by which the printer magnifies or decimates the image in order to fit the image in the image box on film. |
||
Further specifies the type of the interpolation function. Values are defined in Conformance Statement. Only valid for Magnification Type (2010,0060) = CUBIC. |
||
Density of the film areas surrounding and between images on the film. i where i represents the desired density in hundredths of OD (e.g., 150 corresponds with 1.5 OD). |
||
Density of the image box area on the film that contains no image. i where i represents the desired density in hundredths of OD (e.g., 150 corresponds with 1.5 OD). |
||
Minimum density of the images on the film, expressed in hundredths of OD. If Min Density is lower than minimum printer density than Min Density is set to minimum printer density. |
||
Maximum density of the images on the film, expressed in hundredths of OD. If Max Density is higher than maximum printer density than Max Density is set to maximum printer density. |
||
Specifies whether a trim box shall be printed surrounding each image on the film. |
||
Character string that contains either the ID of the printer configuration table that contains a set of values for implementation specific print parameters (e.g., perception LUT related parameters) or one or more configuration data values, encoded as characters. If there are multiple configuration data values encoded in the string, they shall be separated by backslashes. The definition of values shall be contained in the SCP's Conformance Statement. |
||
Luminance of lightbox illuminating a piece of transmissive film, or for the case of reflective media, luminance obtainable from diffuse reflection of the illumination present. Expressed as L0, in candelas per square meter (cd/m2). |
||
For transmissive film, luminance contribution due to reflected ambient light. Expressed as La, in candelas per square meter (cd/m2). |
||
Specifies the resolution at which images in this Film Box are to be printed. |
||
An ICC Profile encoding the transformation of device-dependent color stored pixel values into PCS-Values. See Section C.11.15. |
The standard format subdivides a film into image boxes of equal size. Therefore, the film layout is fully symmetrical, i.e., the arrangement of image boxes on film is left-right and top-bottom symmetric.
DICOM PS3.3 2024d - Information Object Definitions |
---|